<?xml version="1.0" encoding="UTF-8"?>
<!DOCTYPE article PUBLIC "-//NLM//DTD Journal Publishing with OASIS Tables v3.0 20080202//EN" "journalpub-oasis3.dtd">
<article xmlns:xlink="http://www.w3.org/1999/xlink" xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:oasis="http://docs.oasis-open.org/ns/oasis-exchange/table" dtd-version="3.0">
  <front>
    <journal-meta>
<journal-id journal-id-type="publisher">JSSS</journal-id>
<journal-title-group>
<journal-title>Journal of Sensors and Sensor Systems</journal-title>
<abbrev-journal-title abbrev-type="publisher">JSSS</abbrev-journal-title>
<abbrev-journal-title abbrev-type="nlm-ta">J. Sens. Sens. Syst.</abbrev-journal-title>
</journal-title-group>
<issn pub-type="epub">2194-878X</issn>
<publisher><publisher-name>Copernicus Publications</publisher-name>
<publisher-loc>Göttingen, Germany</publisher-loc>
</publisher>
</journal-meta>

    <article-meta>
      <article-id pub-id-type="doi">10.5194/jsss-5-261-2016</article-id><title-group><article-title>Ultrafast imaging Raman spectroscopy of large-area samples without stepwise
scanning</article-title>
      </title-group><?xmltex \runningtitle{Ultrafast imaging Raman spectroscopy of large-area samples without stepwise
scanning}?><?xmltex \runningauthor{E.~Schm\"{a}lzlin~et al.}?>
      <contrib-group>
        <contrib contrib-type="author" corresp="yes" rid="aff1">
          <name><surname>Schmälzlin</surname><given-names>Elmar</given-names></name>
          <email>eschmaelzlin@aip.de</email>
        </contrib>
        <contrib contrib-type="author" corresp="no" rid="aff1">
          <name><surname>Moralejo</surname><given-names>Benito</given-names></name>
          
        </contrib>
        <contrib contrib-type="author" corresp="no" rid="aff1">
          <name><surname>Bodenmüller</surname><given-names>Daniel</given-names></name>
          
        </contrib>
        <contrib contrib-type="author" corresp="no" rid="aff2">
          <name><surname>Darvin</surname><given-names>Maxim E.</given-names></name>
          
        </contrib>
        <contrib contrib-type="author" corresp="no" rid="aff2">
          <name><surname>Thiede</surname><given-names>Gisela</given-names></name>
          
        </contrib>
        <contrib contrib-type="author" corresp="no" rid="aff1">
          <name><surname>Roth</surname><given-names>Martin M.</given-names></name>
          
        </contrib>
        <aff id="aff1"><label>1</label><institution>Leibniz-Institut für Astrophysik Potsdam (AIP), Potsdam, 14482,
Germany</institution>
        </aff>
        <aff id="aff2"><label>2</label><institution>Center of Experimental and Applied Cutaneous Physiology, Department of
Dermatology, Venerology and Allergology, Charité –
Universitätsmedizin Berlin, Berlin, 10117, Germany</institution>
        </aff>
      </contrib-group>
      <author-notes><corresp id="corr1">Elmar Schmälzlin (eschmaelzlin@aip.de)</corresp></author-notes><pub-date><day>13</day><month>July</month><year>2016</year></pub-date>
      
      <volume>5</volume>
      <issue>2</issue>
      <fpage>261</fpage><lpage>271</lpage>
      <history>
        <date date-type="received"><day>9</day><month>February</month><year>2016</year></date>
           <date date-type="rev-recd"><day>20</day><month>May</month><year>2016</year></date>
           <date date-type="accepted"><day>23</day><month>June</month><year>2016</year></date>
      </history>
      <permissions>
<license license-type="open-access">
<license-p>This work is licensed under a Creative Commons Attribution 3.0 Unported License. To view a copy of this license, visit <ext-link ext-link-type="uri" xlink:href="http://creativecommons.org/licenses/by/3.0/">http://creativecommons.org/licenses/by/3.0/</ext-link></license-p>
</license>
</permissions><self-uri xlink:href="https://jsss.copernicus.org/articles/5/261/2016/jsss-5-261-2016.html">This article is available from https://jsss.copernicus.org/articles/5/261/2016/jsss-5-261-2016.html</self-uri>
<self-uri xlink:href="https://jsss.copernicus.org/articles/5/261/2016/jsss-5-261-2016.pdf">The full text article is available as a PDF file from https://jsss.copernicus.org/articles/5/261/2016/jsss-5-261-2016.pdf</self-uri>


      <abstract>
    <p>Step-by-step, time-consuming scanning of the sample is still the
state-of-the-art in imaging Raman spectroscopy. Even for a few 100 image
points the measurement time may add up to minutes or hours. A radical
decrease in measurement time can be achieved by applying multiplex
spectrographs coupled to imaging fiber bundles that are successfully used in
astronomy. For optimal use of the scarce and expensive observation time at
astronomical observatories, special high-performance spectrograph systems
were developed. They are designed for recording thousands of spatially
resolved spectra of a two-dimensional image field within one single exposure.
Transferring this technology to imaging Raman spectroscopy allows a
considerably faster acquisition of chemical maps. Currently, an imaging field
of up to 1 cm<inline-formula><mml:math display="inline"><mml:msup><mml:mi/><mml:mn mathvariant="normal">2</mml:mn></mml:msup></mml:math></inline-formula> can be investigated. For porcine skin the required
measurement time is less than 1 min. For this reason, this technique is of
particular interest for medical diagnostics, e.g., the identification of
potentially cancerous abnormalities of skin tissue.</p>
  </abstract>
    </article-meta>
  </front>
<body>
      

<sec id="Ch1.S1" sec-type="intro">
  <title>Introduction</title>
      <p>When monochromatic light impinges on a sample, the inelastically scattered
light is wavelength shifted (Raman effect). The shift contains information
about the vibration states of the impinged molecules and any crystal
structures. Raman spectra are fingerprints, which allow a contact- and
label-free identification of chemical structures. The underlying physical
effect was already discovered in 1928 by Chandrasekhara Venkata Raman.
Unfortunately, Raman lines are extremely faint (only a fraction of <inline-formula><mml:math display="inline"><mml:mrow><mml:mn mathvariant="normal">1</mml:mn><mml:mo>×</mml:mo><mml:msup><mml:mn>10</mml:mn><mml:mrow><mml:mo>-</mml:mo><mml:mn mathvariant="normal">7</mml:mn></mml:mrow></mml:msup></mml:mrow></mml:math></inline-formula> of the scattered light is Raman scattering). For this reason, the
number of practical applications of Raman spectroscopy was for a long time
rather limited. However, based on considerable progress in laser and detector
technologies, the situation has changed completely. In particular, in the
fields of medical diagnostics and life sciences, Raman microscopes have
become more and more common. In contrast to the traditional infrared
absorption spectroscopy, Raman spectroscopy is not impeded by the infrared
light absorption of water. Raman lines appear relative to the laser
excitation wavelength. When using excitation light at visible wavelengths,
the Raman lines always appear in a range where the absorption of water is
negligibly small. The Raman activity of water is also small. Moreover, in the
majority of cases the spectral positions of water Raman signals are outside
of the region of interest. Thus, Raman spectroscopy works well in aqueous
environments. This turns out to be a key advantage with regard to biological
samples. Reviews of methods and applications of Raman imaging are given in
Opilik et al. (2013) and Stewart et al. (2012). In the field of medical
diagnostics, imaging Raman spectroscopy is a promising technique for
identifying cancerous parts of tissue (Reble et al., 2014; Zhao et al.,
2015). A Raman microscope with a special design for the examination of skin
in vivo on the patient is already commercially available (gen2-SCA Raman
system for in vivo skin analysis. RiverD International B.V., Rotterdam, the
Netherlands, <uri>http://www.riverd.com</uri>, last access 18 May 2016).</p>
      <p>However, like common Raman microscopes, this instrument can only detect Raman
scattering arising from a single spot with a diameter in the range of
micrometers. By contrast, the diagnostics of potentially cancerous skin areas
must be performed in the range of at least some square millimeters. Standard
point-by-point scanning is time-consuming. Even for an image being composed
of some 100 pixels, the measurement time can exceed 1 h. For examination at
a patient or for fast-changing samples, this time span is far too long to be
practical. To accelerate the measuring procedure, various techniques for a
parallel data collection are described (Li et al., 2013; Hagen et al., 2012).
Of particular interest are full-throughput snapshot techniques, also called
multi-channel or integral field spectroscopy (IFS). They work without any
tuning procedures and fundamentally sacrifice no light during the recording procedure.
In the field of astronomy, multi-channel spectroscopy techniques have been
developed since more than 3 decades ago (Vanderriest, 1980) to record images
of faint galaxies that are both spatially and spectrally resolved, and that
are obtained with one single exposure. By avoiding any scanning procedure,
the expensive observation time at large telescopes is thus utilized more
efficiently. A review of the breakthrough obtained with this technique can be
found in Roth (2010). The currently most powerful spectrograph, MUSE
(Multi-Unit Spectroscopic Explorer, Bacon et al., 2004,
<uri>http://www.eso.org/sci/facilities/paranal/instruments/muse.html</uri>, last
access on 18 May 2016), was commissioned at the Very Large Telescope
observatory in Paranal, Chile, in spring 2014.</p>
      <p>MUSE consists of 24 connected spectrograph modules and is capable of
acquiring an image field of the sky with a size of 1 arcmin<inline-formula><mml:math display="inline"><mml:msup><mml:mi/><mml:mn mathvariant="normal">2</mml:mn></mml:msup></mml:math></inline-formula>, a
sampling of <inline-formula><mml:math display="inline"><mml:mrow><mml:mn>0.2</mml:mn><mml:mo>×</mml:mo><mml:mn>0.2</mml:mn></mml:mrow></mml:math></inline-formula> arsec<inline-formula><mml:math display="inline"><mml:msup><mml:mi/><mml:mn mathvariant="normal">2</mml:mn></mml:msup></mml:math></inline-formula>, i.e., a total number of 90 000
spatial elements (“spaxels”), and a depth of 4000 spectral bins within one
single exposure. From every single spaxel the entire spectrum from 465 to
930 nm can be recorded at a spectral resolution of 0.25 nm. As part of a
technology transfer project, a 25th MUSE spectrograph module was procured and
installed in an AIP laboratory (Roth et al., 2012). The aim of the project is
to investigate the suitability of fiber-coupled IFS for the identification of
skin abnormalities including cancerous areas by means of imaging Raman
spectroscopy. With regard to patient friendliness, the measurement times
should be kept as short as possible.</p>
</sec>
<sec id="Ch1.S2">
  <title>Materials and methods</title>
<sec id="Ch1.S2.SS1">
  <title>Fiber bundle-coupled astronomy spectrograph</title>
      <p>The basic idea of IFS is slicing a two-dimensional image into single stripes
and assembling them longitudinally in front of the entrance slit of a
spectrograph (Allington-Smith, 2006). Amongst other techniques, this can be
performed by an image slicer, i.e., a mirror stack followed by a recombining
optic, or by use of a fiber bundle. Fiber bundle-coupled IFS is also called
FAST (fiber array spectral translation, Stewart et al., 2012) or FRIS
(fiber-reformatting imaging spectroscopy, Hagen et al., 2012). Figure 1
describes the concept. At the sample side, the image of the investigated
object, which may also be a Raman or fluorescence image, is focused at the
surface of a two-dimensional fiber array. The fibers collect the signal of
the image. At the spectrograph side, the fibers of the bundle are arranged in
one row forming a pseudo slit for the spectrograph. The signals emerging from
each fiber are collimated, pass through a dispersive element, and finally
form spectrally dispersed light traces on a charge-coupled device (CCD) chip. Data reduction software extracts and
calibrates the CCD raw signal and finally produces a data cube that contains
the entire spectral and spatial information. The setup works in essence as if
every single fiber were connected to an individual single-channel
spectrograph.</p>

      <?xmltex \floatpos{t}?><fig id="Ch1.F1"><caption><p>Function principle of fiber bundle-coupled IFS. The image of the
sample (red and green semi-circles) is focused on the fiber matrix of the
image acquisition head. At the spectrograph side, the fibers are arranged in
a one-dimensional row. After passing a spectral disperser (e.g., a volume
phase holographic grating, VPHG), every fiber produces an individual
intensity- and spectrum-dependent light trace at the CCD chip. The raw signal
of the CCD is evaluated by software and converted into a data cube that
contains the entire spectral and spatial information.</p></caption>
          <?xmltex \igopts{width=236.157874pt}?><graphic xlink:href="https://jsss.copernicus.org/articles/5/261/2016/jsss-5-261-2016-f01.png"/>

        </fig>

      <p>The employed astronomy spectrograph and the fiber bundle were presented
previously (Schmälzlin et al., 2014). In brief: the MUSE spectrograph
unit is a high-performance fully refractive optical system that covers a
spectral range from 465 to 930 nm with about 0.25 nm resolution. The size
of one unit including the CCD camera housing is approximately
0.6 m <inline-formula><mml:math display="inline"><mml:mo>×</mml:mo></mml:math></inline-formula> 1.9 m <inline-formula><mml:math display="inline"><mml:mo>×</mml:mo></mml:math></inline-formula> 0.6 m. It comprises a volume phase
holographic grating (VPHG) and a liquid-nitrogen cooled 4 K <inline-formula><mml:math display="inline"><mml:mo>×</mml:mo></mml:math></inline-formula> 4 K
pixel CCD detector system. The VPGH, consisting of a phase-modulating
dichromatic gelatin layer that is sealed between two glass plates, shows up
to 80 % diffraction efficiency and minimally scattered light over the
whole spectral range. <italic>p3d</italic>, an open-source data reduction and
analysis software (<uri>http://p3d.sourceforge.net/</uri>, last access 18 May
2016) that was initially developed for astronomical purposes (Sandin at al.,
2010), has been customized for that project to calculate the spectra from the
raw CCD image and to correlate them with the source position at the
two-dimensional fiber array. Finally, a data cube is generated, which
contains all spectral and spatial information for further data processing.
The fiber bundle contains 400 step index multimode fibers (110 <inline-formula><mml:math display="inline"><mml:mi mathvariant="normal">µ</mml:mi></mml:math></inline-formula>m
core), forming a 20 <inline-formula><mml:math display="inline"><mml:mo>×</mml:mo></mml:math></inline-formula> 20 matrix at the image acquisition side
(Fig. 2). Hence, a spectrally resolved square image composed of 400 pixels
can be recorded within one single exposure. The pitch between the fiber core
centers is 0.5 mm, which is an acceptable tradeoff between resolution and
crosstalk with regard to the scattering properties of skin samples.</p>

      <?xmltex \floatpos{t}?><fig id="Ch1.F2"><caption><p>(Reprinted from Schmälzlin et al., 2015): image acquisition head
with a <inline-formula><mml:math display="inline"><mml:mrow><mml:mn>20</mml:mn><mml:mo>×</mml:mo><mml:mn>20</mml:mn></mml:mrow></mml:math></inline-formula> fiber matrix. For demonstration purposes of this picture
the fibers are partially illuminated at the distal end of the fiber bundle.</p></caption>
          <?xmltex \igopts{width=236.157874pt}?><graphic xlink:href="https://jsss.copernicus.org/articles/5/261/2016/jsss-5-261-2016-f02.jpg"/>

        </fig>

</sec>
<sec id="Ch1.S2.SS2">
  <title>Optics for the acquisition of large-area Raman images</title>
      <p>The use of fiber bundle coupled IFS for imaging Raman spectroscopy was
already reported (Ma and Ben-Amotz, 1997; Okuno and Hamaguchi, 2010; Roth et
al., 2012; Stewart et al., 2012; Schmälzlin et al., 2014, 2015;
Brückner et al., 2015). However, previous setups were designed to examine
samples with microscopic dimension, e.g., cells or tiny crystal fragments.
Here, we present a setup that is capable of acquiring Raman images from
samples with a size of up to 1 cm<inline-formula><mml:math display="inline"><mml:msup><mml:mi/><mml:mn mathvariant="normal">2</mml:mn></mml:msup></mml:math></inline-formula>, which is the typical size of
potentially cancerous skin patches (Reble et al., 2014). Two optic setups
were developed. Acquisition optics 1 (AO1) allows recording of Raman images
from square surfaces with 3.6 mm edge length. Acquisition optics 2 (AO2) is
even suitable for a square area with 1 cm edge length. In order to receive
large-area Raman images within one single exposure and without scanning, it
is favorable to illuminate the whole area of interest as uniformly as
possible with the excitation laser. This was performed by feeding the laser
light with a square core fiber (F&amp;T Fibers and Technology GmbH, Germany,
VIS-IR, square core with 600 <inline-formula><mml:math display="inline"><mml:mi mathvariant="normal">µ</mml:mi></mml:math></inline-formula>m edge length). Highly multimode
square core fibers, which have been designed initially for laser material
processing, shape the incoming Gaussian-like beam to a top-hat intensity
distribution (Brückner et al., 2015). Figure 3 shows the resulting square
and uniform excitation light spot generated by AO1 at a millimeter-scaled
paper. The spot is somewhat larger than the actual image field of the sample
to exclude the effect of possible intensity inequalities at the spot's
borders.</p>

      <?xmltex \floatpos{t}?><fig id="Ch1.F3"><caption><p>Excitation spot generated with AO1 at the sample plane. The
pitch of the scaled paper is 1 mm.</p></caption>
          <?xmltex \igopts{width=236.157874pt}?><graphic xlink:href="https://jsss.copernicus.org/articles/5/261/2016/jsss-5-261-2016-f03.jpg"/>

        </fig>

      <p>Figure 4 shows a schematic drawing and a photo of AO1. The excitation laser
light (785 nm) is fed via the square core glass fiber, passes a collimator
and a 785 nm clean-up filter (Edmund Optics) and is finally guided to the
sample by use of a 45<inline-formula><mml:math display="inline"><mml:msup><mml:mi/><mml:mo>∘</mml:mo></mml:msup></mml:math></inline-formula> dichroic mirror (Semrock, Di02-R785-25x36)
that is transparent for Stokes-shifted Raman signals. An achromatic doublet
(<inline-formula><mml:math display="inline"><mml:mrow><mml:mi>f</mml:mi><mml:mo>=</mml:mo><mml:mn>50</mml:mn></mml:mrow></mml:math></inline-formula> mm) focuses the laser light on the sample and collects the Raman
scattering. A further dichroic mirror (755 nm long pass, Semrock, USA,
FF757-Di01-25x36) takes visible light from the sample and guides it to a
camera that is used to position and focus the sample. A 785 nm long pass
filter (Semrock, USA, BLP01-785R-25) blocks the excitation laser light and
the Rayleigh scattering. A further achromatic doublet (<inline-formula><mml:math display="inline"><mml:mrow><mml:mi>f</mml:mi><mml:mo>=</mml:mo><mml:mn>125</mml:mn></mml:mrow></mml:math></inline-formula> mm,
NA <inline-formula><mml:math display="inline"><mml:mrow><mml:mo>≈</mml:mo><mml:mn>0.1</mml:mn></mml:mrow></mml:math></inline-formula>) focuses the image of the sample, strictly speaking the
arising Raman signals, onto the surface of the fiber matrix. For convenient
positioning of the sample, an XY stage linked to a micrometer lifting table
was used. The measurement head was mounted below the opening of the stage
plate.</p>

      <?xmltex \floatpos{t}?><fig id="Ch1.F4"><caption><p>Design and photo of image acquisition optic AO1 suitable for a
square image field with 3.6 mm side length.</p></caption>
          <?xmltex \igopts{width=236.157874pt}?><graphic xlink:href="https://jsss.copernicus.org/articles/5/261/2016/jsss-5-261-2016-f04.jpg"/>

        </fig>

      <p>A weak point in the design of AO1 is the poor coupling efficiency at the side
of the fiber matrix. The fiber matrix has an area of 100 mm<inline-formula><mml:math display="inline"><mml:msup><mml:mi/><mml:mn mathvariant="normal">2</mml:mn></mml:msup></mml:math></inline-formula>; however,
only 4 mm<inline-formula><mml:math display="inline"><mml:msup><mml:mi/><mml:mn mathvariant="normal">2</mml:mn></mml:msup></mml:math></inline-formula> consist of fiber cores. To improve the situation, a further
optic unit (AO2) with a first microlens array (MLA; Advanced Microoptic
Systems, Germany, APO-Q-P500(D500)-AR0.58) in front of the fiber matrix was
designed. The MLA consists of <inline-formula><mml:math display="inline"><mml:mrow><mml:mn>24</mml:mn><mml:mo>×</mml:mo><mml:mn>24</mml:mn></mml:mrow></mml:math></inline-formula> photolithographical-manufactured,
fused silica microlenses with 1.3 mm focus lengths, 0.5 mm pitch, and
0.5 mm lens diameters. The clear aperture of each lens is 0.45 mm,
resulting in a fill factor of 65 %. The flat side of the MLA points
towards the fiber matrix; <inline-formula><mml:math display="inline"><mml:mrow><mml:mn>20</mml:mn><mml:mo>×</mml:mo><mml:mn>20</mml:mn></mml:mrow></mml:math></inline-formula> lenses are actively used to focus the
signal into the fiber cores and to reduce losses at blind areas of the fiber
matrix. Figure 5 shows a schematic drawing of AO2. A second MLA of the same
type was used to increase the image field at the sample side to
10 mm <inline-formula><mml:math display="inline"><mml:mo>×</mml:mo></mml:math></inline-formula> 10 mm, which was an important objective for a future skin
cancer detection application.</p>

      <?xmltex \floatpos{t}?><fig id="Ch1.F5"><caption><p>Design of the microlens-based image acquisition optic AO2
suitable for a square image field with 1 cm side length.</p></caption>
          <?xmltex \igopts{width=236.157874pt}?><graphic xlink:href="https://jsss.copernicus.org/articles/5/261/2016/jsss-5-261-2016-f05.png"/>

        </fig>

      <p>The flat side of the second MLA points towards the sample. Figure 6 shows
the excitation laser spots at a millimeter scaled paper. A pair of relay
lenses guides the Raman signal from the second to the first MLA. Although
the realization of AO2 is still preliminary, first test measurements with
AO2 could be performed successfully.</p>

      <?xmltex \floatpos{t}?><fig id="Ch1.F6"><caption><p>Spots of the excitation laser at about 0.8 mm distance from the
flat surface of the MLA. The pitch of the scaled paper is 1 mm. The slanted
shadings at the lower left and at the upper right are due to distance
holders.</p></caption>
          <?xmltex \igopts{width=236.157874pt}?><graphic xlink:href="https://jsss.copernicus.org/articles/5/261/2016/jsss-5-261-2016-f06.jpg"/>

        </fig>

      <p>For biological samples, a vessel with a calcium fluoride bottom is available
(Fig. 8). Raman grade calcium fluoride is a suitable slide material since it
generates a low background signal.</p>
</sec>
<sec id="Ch1.S2.SS3">
  <title>Laser source</title>
      <p>As a light source, a fiber-coupled dual laser light source (Newport, USA,
LS-2-7878-FC) that contains two switchable diode lasers with 784.5 and
785.5 nm was used. The maximum output power of each laser is 500 mW;
however, the laser power was normally aligned to 230 mW measured in the
sample plane of AO1. This corresponds to a power density of 14 mW mm<inline-formula><mml:math display="inline"><mml:msup><mml:mi/><mml:mrow><mml:mo>-</mml:mo><mml:mn mathvariant="normal">2</mml:mn></mml:mrow></mml:msup></mml:math></inline-formula>
in the case of a uniformly illuminated square with 4.1 mm side length. An
estimation of the power density at the sample plane of AO2 is done as
follows: at a 100 mm<inline-formula><mml:math display="inline"><mml:msup><mml:mi/><mml:mn mathvariant="normal">2</mml:mn></mml:msup></mml:math></inline-formula> sample field, i.e., in the field of 400 spots, a
total laser power of 243 mW is measured. According to calculations (Zemax
LLC, USA, Opticstudio), the predicted diameter of one single laser spot under
perfect alignment conditions is 0.01 mm. The actual value is roughly
0.05 mm. Thus, 400 spots illuminate a total area of 0.8 mm<inline-formula><mml:math display="inline"><mml:msup><mml:mi/><mml:mn mathvariant="normal">2</mml:mn></mml:msup></mml:math></inline-formula>,
resulting in about 300 mW mm<inline-formula><mml:math display="inline"><mml:msup><mml:mi/><mml:mrow><mml:mo>-</mml:mo><mml:mn mathvariant="normal">2</mml:mn></mml:mrow></mml:msup></mml:math></inline-formula> power density within a spot. Taking
into account the 65 % fill factor of the MLA, i.e., excluding the light
amount passing between the individual lenses, the power density is reduced to
195 mW mm<inline-formula><mml:math display="inline"><mml:msup><mml:mi/><mml:mrow><mml:mo>-</mml:mo><mml:mn mathvariant="normal">2</mml:mn></mml:mrow></mml:msup></mml:math></inline-formula>. It is worth noting that common single channel Raman
microscopes work with far higher power densities. For instance, the
microscope used in Reble et al. (2014) applies 70 mW to a circular spot with
0.16 mm diameter, which corresponds to 3.5 W mm<inline-formula><mml:math display="inline"><mml:msup><mml:mi/><mml:mrow><mml:mo>-</mml:mo><mml:mn mathvariant="normal">2</mml:mn></mml:mrow></mml:msup></mml:math></inline-formula>.</p>
</sec>
<sec id="Ch1.S2.SS4">
  <title>SERDS measurements</title>
      <p>The elimination of background fluorescence is a challenge in the field of
Raman spectroscopy. In principle, an increase of the excitation wavelength
helps to reduce background fluorescence. However, Raman scattering intensity
also decreases, since it depends on the fourth power of the excitation
frequency. The 785 nm excitation wavelength is often used for organic
samples, since it is an acceptable tradeoff between fluorescence and Raman
intensity. Furthermore, a wide range of detectors, lasers, and optical
filters is available for this wavelength. Nevertheless, the amount of
fluorescence may exceed the Raman signal intensity by several orders of
magnitude. Various methods are described to remove the fluorescence
background (Wei et al., 2015). Most common are post-processing computational
methods, e.g., polynomial fitting and subsequent subtracting of the
background. A more advanced background removal method is shifted excitation
Raman difference spectroscopy (SERDS; Sowoidnich and Kronfeldt, 2012). It is
based on the fact that Raman signals always appear relative to the excitation
laser wavelength, but fluorescence emission occurs at fixed wavelengths. For
SERDS, the Raman spectra of a sample are recorded twice with two slightly
different excitation wavelengths (here 784.5 and 785.5 nm, respectively)
followed by the application of an appropriate subtraction and transformation
procedure. Finally, background-free Raman spectra are reconstructed. The used
algorithm is shown in Schmidt et al. (2014).</p>
</sec>
<sec id="Ch1.S2.SS5">
  <title>Samples</title>
      <p>The capability of the described setups was demonstrated with the following
samples: the Aspirin Effect pain reliever, a piece of silicon, a EUR 10 bank
bill, the cut face of a porcine ear, porcine skin, and a piece of brown sugar
dissolving in water.</p>
</sec>
</sec>
<sec id="Ch1.S3">
  <title>Result and discussions</title>
<sec id="Ch1.S3.SS1">
  <title>Aspirin Effect</title>
      <p>Aspirin Effect (Bayer Bitterfeld GmbH, Germany) is a pain reliever tablet
consisting of rod-like granules with about 1 mm length and relatively
unshaped fragments. According to the package information sheet, it contains
the active agent acetylsalicylic acid, D mannitol, sodium hydrogen carbonate,
sodium dihydrogen citrate, ascorbic acid, citric acid, flavors, and the
aspartame sweetener. A small spatula tip of a granule was put on a 1 mm
thick calcium fluoride slide and placed on the XY stage over AO1. The Raman
spectra were recorded at 784.5 and 785.5 nm excitation wavelength. The
measurement times were 1 s for each wavelength. The read-out time of the
full CCD chip was chosen to be 10 s, which is the shortest possible without
binning. Thus, the effective measurement time for a complete data set was
22 s. For all further experiments a 10 s read-out time was chosen as well.
Figure 7 shows the camera image, the SERDS spectra at two picked positions
and the Raman image. The Raman image illustrates the intensity distribution
of the Raman signal at 1606 cm<inline-formula><mml:math display="inline"><mml:msup><mml:mi/><mml:mrow><mml:mo>-</mml:mo><mml:mn mathvariant="normal">1</mml:mn></mml:mrow></mml:msup></mml:math></inline-formula> that is characteristic of
acetylsalicylic acid. Obviously, the acetylsalicylic acid agent is present in
the form of the fine powder. The rod-like granules contain no active
agent.<?xmltex \hack{\newpage}?></p>

      <?xmltex \floatpos{t}?><fig id="Ch1.F7"><caption><p>Granules and powder of the Aspirin Effect pain reliever. Upper left:
camera picture. Upper right: pseudo color Raman image of the intensity at
1606 cm<inline-formula><mml:math display="inline"><mml:msup><mml:mi/><mml:mrow><mml:mo>-</mml:mo><mml:mn mathvariant="normal">1</mml:mn></mml:mrow></mml:msup></mml:math></inline-formula>, which corresponds to the distribution of the acetylsalicylic
acid agent. Below: corresponding SERDS spectra at positions
<?xmltex \hack{\protect}?><?xmltex \igopts{width=5.690551pt}?><inline-graphic xlink:href="https://jsss.copernicus.org/articles/5/261/2016/jsss-5-261-2016-g02.pdf"/> and
<?xmltex \hack{\protect}?><?xmltex \igopts{width=5.690551pt}?><inline-graphic xlink:href="https://jsss.copernicus.org/articles/5/261/2016/jsss-5-261-2016-g01.pdf"/> (1 s
recording time per laser).</p></caption>
          <?xmltex \igopts{width=236.157874pt}?><graphic xlink:href="https://jsss.copernicus.org/articles/5/261/2016/jsss-5-261-2016-f07.png"/>

        </fig>

</sec>
<sec id="Ch1.S3.SS2">
  <title>Silicon</title>
      <p>A small silicon plate with 0.66 mm thickness is used for testing AO2 and a
comparison with AO1. Pure silicon shows one single Raman signal around
520 cm<inline-formula><mml:math display="inline"><mml:msup><mml:mi/><mml:mrow><mml:mo>-</mml:mo><mml:mn mathvariant="normal">1</mml:mn></mml:mrow></mml:msup></mml:math></inline-formula> (Parker, 1967). To determine the optimal distance from the
surface of the microlens array, the silicon plate was put in a tilted
position on the top of AO2. Figure 8 left shows the silicon plate in front of
the microlens array. At the upper right, the silicon plate touches the head
of a turned screw. At this point of support the distance from the MLA surface
is <inline-formula><mml:math display="inline"><mml:mrow><mml:msub><mml:mi>d</mml:mi><mml:mn mathvariant="normal">1</mml:mn></mml:msub><mml:mo>=</mml:mo><mml:mn>1.9</mml:mn></mml:mrow></mml:math></inline-formula> mm. At the lower left, the silicon plate touches the corner
MLA surface (<inline-formula><mml:math display="inline"><mml:mrow><mml:msub><mml:mi>d</mml:mi><mml:mn mathvariant="normal">0</mml:mn></mml:msub><mml:mo>=</mml:mo><mml:mn mathvariant="normal">0</mml:mn></mml:mrow></mml:math></inline-formula> mm). Figure 8 right shows the intensity distribution
of the silicon Raman peak at 522 cm<inline-formula><mml:math display="inline"><mml:msup><mml:mi/><mml:mrow><mml:mo>-</mml:mo><mml:mn mathvariant="normal">1</mml:mn></mml:mrow></mml:msup></mml:math></inline-formula>. The Raman spectra are recorded
at 784.5 nm for 20 s. The rolling circle method (Brandt et al., 2006) was
applied to remove fluorescence background arising from the optics. Obviously,
pixels located at the borders are effected by vignetting. The microlens rows
at the right and the lower borders are close to the contact area of the
holder plate. The given cutting lines of the MLA and the limited adjustment
possibilities of AO2 did not allow for a more favorable position.
Furthermore, the visual impression of the image of the laser spots (Fig. 8)
suggests that the excitation intensities are somewhat lower at the border
lines. Apart from this limitation the shape of the silicon piece is
reproduced very well. Around the spatial position <inline-formula><mml:math display="inline"><mml:mrow><mml:mo>(</mml:mo><mml:mi>X</mml:mi><mml:mo>/</mml:mo><mml:mi>Y</mml:mi><mml:mo>)</mml:mo><mml:mo>=</mml:mo><mml:mo>(</mml:mo><mml:mn mathvariant="normal">6</mml:mn><mml:mo>/</mml:mo><mml:mn mathvariant="normal">8</mml:mn><mml:mo>)</mml:mo></mml:mrow></mml:math></inline-formula> there is an
obvious distance-dependent signal maximum of about 3000 counts per pixel. The
maximum position was transferred to camera picture and the distances to the
two support points were measured. Triangulation yields a distance of
<inline-formula><mml:math display="inline"><mml:mrow><mml:msub><mml:mi>d</mml:mi><mml:mtext>focus</mml:mtext></mml:msub><mml:mo>=</mml:mo><mml:mn>0.8</mml:mn></mml:mrow></mml:math></inline-formula> mm that points to the highest signal intensity. The
experiment was repeated by using Raman grade calcium fluoride windows with
thicknesses of 0.6, 0.7, 0.8 and 1.0 mm, respectively. The windows were
placed as distance holders between the MLA and the silicon plate. The maximum
average Raman signal intensity of the silicon piece was detected when using a
0.8 mm window. Computer simulations of the AO2 setup (Zemax LLC, USA,
Opticstudio) predict an optimal distance of 0.6 mm in the case of an air gap
between MLA surface and sample and 1.0 mm using a calcium fluoride window as
spacer. The differences from the experiment values are likely due to
adjustment deficits of AO2. Simulations show that the focus at the sample
side depends much on the fine adjustment of the distal MLA. However, a
precise alignment in the micrometer range and an accurate control of the
actual positions of the optical components cannot be realized with the
current AO2 setup.</p>

      <?xmltex \floatpos{t}?><fig id="Ch1.F8"><caption><p>A tilted silicon piece is put in front of the sample-side
MLA of AO2 (left). The distant-dependent distribution of the silicon's Raman
signal is plotted in pseudo colors (right). Triangulation yields an optimal
sample distance <inline-formula><mml:math display="inline"><mml:mrow><mml:msub><mml:mi>d</mml:mi><mml:mtext>focus</mml:mtext></mml:msub><mml:mo>=</mml:mo><mml:mn>0.8</mml:mn></mml:mrow></mml:math></inline-formula> mm, showing the highest Raman intensity.</p></caption>
          <?xmltex \igopts{width=236.157874pt}?><graphic xlink:href="https://jsss.copernicus.org/articles/5/261/2016/jsss-5-261-2016-f08.png"/>

        </fig>

      <p>For comparison, the silicon plate was also examined with AO1. The silicon
plate was positioned at the focal plane of AO1 and the Raman spectra were
recorded with an excitation wavelength of 784.5 nm and an exposure time of
20 s. The settings of the diode laser unit were the same as for the
measurement with AO2. The average intensity of the Raman signal achieved with
AO1 was 1000 counts per pixel. Thus, the MLA-based AO2 provides 3 times more
counts per pixel than AO1. To estimate the amount of improvement, the
following considerations were made: the image field sizes captured with
one single fiber of the array are
13 mm<inline-formula><mml:math display="inline"><mml:msup><mml:mi/><mml:mn mathvariant="normal">2</mml:mn></mml:msup></mml:math></inline-formula>/<inline-formula><mml:math display="inline"><mml:mrow><mml:mn>400</mml:mn><mml:mo>×</mml:mo><mml:mn>0.04</mml:mn><mml:mo>=</mml:mo><mml:mn>1.3</mml:mn><mml:mo>×</mml:mo><mml:msup><mml:mn>10</mml:mn><mml:mrow><mml:mo>-</mml:mo><mml:mn mathvariant="normal">3</mml:mn></mml:mrow></mml:msup></mml:mrow></mml:math></inline-formula> mm<inline-formula><mml:math display="inline"><mml:msup><mml:mi/><mml:mn mathvariant="normal">2</mml:mn></mml:msup></mml:math></inline-formula> in the case of AO1
(taking into account that without an MLA only 4 % of the projected signal
couples into the fibers of the matrix; see Sect. 2.3) and 100 mm<inline-formula><mml:math display="inline"><mml:msup><mml:mi/><mml:mn mathvariant="normal">2</mml:mn></mml:msup></mml:math></inline-formula>/<inline-formula><mml:math display="inline"><mml:mrow><mml:mn>400</mml:mn><mml:mo>=</mml:mo><mml:mn>0.25</mml:mn></mml:mrow></mml:math></inline-formula> mm<inline-formula><mml:math display="inline"><mml:msup><mml:mi/><mml:mn mathvariant="normal">2</mml:mn></mml:msup></mml:math></inline-formula> in the case of AO2 (assuming ideal MLAs). Within the
13 mm<inline-formula><mml:math display="inline"><mml:msup><mml:mi/><mml:mn mathvariant="normal">2</mml:mn></mml:msup></mml:math></inline-formula> image field of AO1, an excitation laser power of 295 mW was
measured; i.e., the excitation power per image field of one single fiber of
AO1 is 295 mW/13 mm<inline-formula><mml:math display="inline"><mml:mrow><mml:msup><mml:mi/><mml:mn mathvariant="normal">2</mml:mn></mml:msup><mml:mo>×</mml:mo><mml:mn>1.3</mml:mn><mml:mo>×</mml:mo><mml:msup><mml:mn>10</mml:mn><mml:mrow><mml:mo>-</mml:mo><mml:mn mathvariant="normal">3</mml:mn></mml:mrow></mml:msup></mml:mrow></mml:math></inline-formula> mm<inline-formula><mml:math display="inline"><mml:mrow><mml:msup><mml:mi/><mml:mn mathvariant="normal">2</mml:mn></mml:msup><mml:mo>=</mml:mo><mml:mn>0.030</mml:mn></mml:mrow></mml:math></inline-formula> mW. For
AO2, an excitation laser power of 243 mW at an area of 100 mm<inline-formula><mml:math display="inline"><mml:msup><mml:mi/><mml:mn mathvariant="normal">2</mml:mn></mml:msup></mml:math></inline-formula> was
measured, leading to an excitation power per fiber of 243 mW/<inline-formula><mml:math display="inline"><mml:mrow><mml:mn>400</mml:mn><mml:mo>=</mml:mo><mml:mn>0.61</mml:mn></mml:mrow></mml:math></inline-formula> mW.
Consequently, an ideal AO2 should provide twentyfold more Raman signal per
pixel than AO1, meaning that if 1000 counts per pixel are achieved with AO1,
an ideal AO2 should provide 20 000 counts per pixel with the same sample in
focus. However, the real AO2 shows imperfections that can be quantified.
Intermediate space between the single lenses leads to an MLA fill factor of
65 %, reducing the effective power of the excitation light and therefore
decreasing the number of expectable counts per pixel to 20 000 <inline-formula><mml:math display="inline"><mml:mrow><mml:mo>×</mml:mo><mml:mn>0.65</mml:mn><mml:mo>=</mml:mo></mml:mrow></mml:math></inline-formula> 13 000. It is worth noting that the Raman signal of the silicon only
arises from the focus points of the microlenses, i.e., in the places of the
spots of the excitation laser. Thus, the fill factor should not affect the
collection of the Raman signal, at least not in the case of non-scattering
samples like silicon.</p>

      <?xmltex \floatpos{t}?><fig id="Ch1.F9"><caption><p>Part of the letter “E” printed on a EUR 10 bill. Upper left:
camera picture. Upper right: pseudo color Raman image of the intensity at
1576 cm<inline-formula><mml:math display="inline"><mml:msup><mml:mi/><mml:mrow><mml:mo>-</mml:mo><mml:mn mathvariant="normal">1</mml:mn></mml:mrow></mml:msup></mml:math></inline-formula>, which corresponds to the distribution of the printing ink of
the “E”. Below: corresponding SERDS spectra at positions
<?xmltex \hack{\protect}?><?xmltex \igopts{width=5.690551pt}?><inline-graphic xlink:href="https://jsss.copernicus.org/articles/5/261/2016/jsss-5-261-2016-g02.pdf"/> and
<?xmltex \hack{\protect}?><?xmltex \igopts{width=5.690551pt}?><inline-graphic xlink:href="https://jsss.copernicus.org/articles/5/261/2016/jsss-5-261-2016-g01.pdf"/> (10 s
recording time per laser). The measurements were performed with AO1.</p></caption>
          <?xmltex \igopts{width=236.157874pt}?><graphic xlink:href="https://jsss.copernicus.org/articles/5/261/2016/jsss-5-261-2016-f09.png"/>

        </fig>

      <p>Furthermore, the two MLAs do not have antireflection
coatings. Back reflections at four surfaces reduce the expectable counts per
pixel to 13 000 <inline-formula><mml:math display="inline"><mml:mrow><mml:mo>×</mml:mo><mml:mo>(</mml:mo><mml:mn>0.96</mml:mn><mml:msup><mml:mo>)</mml:mo><mml:mn mathvariant="normal">4</mml:mn></mml:msup><mml:mo>=</mml:mo></mml:mrow></mml:math></inline-formula> 11 000. The real AO2 provides 3000
counts per pixel. This implies that in theory AO2 could be approximately even
4 times better. The difference is most likely due to the limited imaging
qualities of the MLAs and the suboptimal alignment of AO2. It should be
emphasized that the current version of AO2 is still a laboratory sample, the
optical components of which are difficult to align in a micrometer range.</p>
</sec>
<sec id="Ch1.S3.SS3">
  <title>EUR 10 bank bill</title>
      <p>A EUR 10 bank bill was placed on a 1 mm thick calcium fluoride slide and
fixed on the XY stage over AO1. A part of the letter “E” of the inscription
“EURO” on the front side is focused. Figure 9 shows the camera image, the
Raman image, and the SERDS spectra at two positions. One position
<?xmltex \hack{\protect}?><?xmltex \igopts{width=5.690551pt}?><inline-graphic xlink:href="https://jsss.copernicus.org/articles/5/261/2016/jsss-5-261-2016-g01.pdf"/> was picked
within, the other
<?xmltex \hack{\protect}?><?xmltex \igopts{width=5.690551pt}?><inline-graphic xlink:href="https://jsss.copernicus.org/articles/5/261/2016/jsss-5-261-2016-g02.pdf"/> outside of
the “E”. The intensity of the line at 1576 cm<inline-formula><mml:math display="inline"><mml:msup><mml:mi/><mml:mrow><mml:mo>-</mml:mo><mml:mn mathvariant="normal">1</mml:mn></mml:mrow></mml:msup></mml:math></inline-formula> seems to be
characteristic of the printing ink of the “E” and was used to generate the
pseudo color Raman image.</p>

      <?xmltex \floatpos{t}?><fig id="Ch1.F10"><caption><p>Comparison of AO1 and the MLA-based AO2 by recording Raman images of
the lettering “EURO” printed on a EUR 10 bill. Upper left: pseudo color
Raman image of a part of the “E” achieved with AO1. Upper right: pseudo
color Raman image of the letters “EU” achieved with AO2. Both Raman images
are based on the intensity at 1576 cm<inline-formula><mml:math display="inline"><mml:msup><mml:mi/><mml:mrow><mml:mo>-</mml:mo><mml:mn mathvariant="normal">1</mml:mn></mml:mrow></mml:msup></mml:math></inline-formula>, which corresponds to the
distribution of the printing ink. Below: average Raman spectra received from
six pixels with AO1 (blue curve) and AO2 (red curve), respectively. The
positions of the binned pixels are outlined with squares in the pseudo color
images.</p></caption>
          <?xmltex \igopts{width=236.157874pt}?><graphic xlink:href="https://jsss.copernicus.org/articles/5/261/2016/jsss-5-261-2016-f10.png"/>

        </fig>

      <p>For comparison, the measurement was also performed with MLA-based AO2. The
part of the bill with the writing “EURO” was placed over the sample-side
MLA. A Raman grade calcium fluoride window was used as a spacer. Its
thickness was 0.7 mm; thus, the focus layer of the excitation light was just
outside the window body. The measurement was performed with only one laser
(784.5 cm<inline-formula><mml:math display="inline"><mml:mrow><mml:msup><mml:mi/><mml:mrow><mml:mo>-</mml:mo><mml:mn mathvariant="normal">1</mml:mn></mml:mrow></mml:msup><mml:mo>)</mml:mo></mml:mrow></mml:math></inline-formula>. Hence, no SERDS spectra were evaluated. The fluorescence
background was removed by means of the rolling circle method (Brandt et al.,
2006). The Raman spectra of the letter “E” received with AO1 at
784.5 cm<inline-formula><mml:math display="inline"><mml:msup><mml:mi/><mml:mrow><mml:mo>-</mml:mo><mml:mn mathvariant="normal">1</mml:mn></mml:mrow></mml:msup></mml:math></inline-formula> previously were likewise background-corrected with the
rolling circle method. The measurement time was 10 s in both cases. The
power output of the laser was not altered when changing from AO1 to AO2.
Figure 10 shows the results. The pseudo color Raman images show the letters
“EU” recorded with AO2 and the part of the “E” recorded with AO1 related
to the intensity of the highest Raman peak of the printing ink at
1576 cm<inline-formula><mml:math display="inline"><mml:msup><mml:mi/><mml:mrow><mml:mo>-</mml:mo><mml:mn mathvariant="normal">1</mml:mn></mml:mrow></mml:msup></mml:math></inline-formula>. Applying the SERDS algorithm leads to arbitrary units,
whereas the rolling circle method retains “counts” as a unit. Thus, the
intensity scales in Figs. 9 and 10 are different. The white square at
<inline-formula><mml:math display="inline"><mml:mrow><mml:mo>(</mml:mo><mml:mn>18</mml:mn><mml:mo>/</mml:mo><mml:mn mathvariant="normal">3</mml:mn><mml:mo>)</mml:mo></mml:mrow></mml:math></inline-formula> is due to a broken fiber of the bundle. The Fig. 10 bottom shows the
corresponding Raman spectra. The blue curve corresponds to AO1, the red one
to AO2. The curves are the averages received from the Raman spectra of,
respectively, six individual pixels. The average was taken to reduce the
noise. The used pixels marked by rectangles are plotted into the pseudo color
images.</p>
      <p>AO2 provides about 10 % higher intensity per pixel than AO1. Indeed, the
MLAs of AO2 improve the signal strength. However, the gain is much lower than
for the silicon plate (see Sect. 3.2). This is likely due to the
comparatively high scattering of the paper. The scattering of the paper
increases the diameters of the excitation laser spots (Fig. 6). Due to the
65 % fill factor of the MLA, Raman signals arising from the borders of
the laser spots that are increased by scattering samples are not collimated.
Furthermore, using a calcium fluoride window affects the light path and thus
the alignment conditions of all optical components. Possibly, the thin bill
was not exactly in the focus plane.</p>

      <?xmltex \floatpos{t}?><fig id="Ch1.F11"><caption><p>Porcine ear with the cut face down at the calcium fluoride
plate of the Raman sample chamber.</p></caption>
          <?xmltex \igopts{width=236.157874pt}?><graphic xlink:href="https://jsss.copernicus.org/articles/5/261/2016/jsss-5-261-2016-f11.jpg"/>

        </fig>

      <?xmltex \floatpos{t}?><fig id="Ch1.F12"><caption><p>Cut face of a porcine ear. Upper left: camera picture. Upper right:
pseudo color Raman image of the intensity at 1303 cm<inline-formula><mml:math display="inline"><mml:msup><mml:mi/><mml:mrow><mml:mo>-</mml:mo><mml:mn mathvariant="normal">1</mml:mn></mml:mrow></mml:msup></mml:math></inline-formula> (lipid band).
Below: corresponding SERDS spectra at positions
<?xmltex \hack{\protect}?><?xmltex \igopts{width=5.690551pt}?><inline-graphic xlink:href="https://jsss.copernicus.org/articles/5/261/2016/jsss-5-261-2016-g02.pdf"/> and
<?xmltex \hack{\protect}?><?xmltex \igopts{width=5.690551pt}?><inline-graphic xlink:href="https://jsss.copernicus.org/articles/5/261/2016/jsss-5-261-2016-g01.pdf"/> (60 s
recording time per laser).</p></caption>
          <?xmltex \igopts{width=236.157874pt}?><graphic xlink:href="https://jsss.copernicus.org/articles/5/261/2016/jsss-5-261-2016-f12.png"/>

        </fig>

      <?xmltex \floatpos{t}?><fig id="Ch1.F13"><caption><p>Porcine skin. Upper left: camera picture. Upper right: pseudo color
Raman image of the intensity at 1455 cm<inline-formula><mml:math display="inline"><mml:msup><mml:mi/><mml:mrow><mml:mo>-</mml:mo><mml:mn mathvariant="normal">1</mml:mn></mml:mrow></mml:msup></mml:math></inline-formula>. Below: corresponding SERDS
spectra at positions
<?xmltex \hack{\protect}?><?xmltex \igopts{width=5.690551pt}?><inline-graphic xlink:href="https://jsss.copernicus.org/articles/5/261/2016/jsss-5-261-2016-g02.pdf"/> and
<?xmltex \hack{\protect}?><?xmltex \igopts{width=5.690551pt}?><inline-graphic xlink:href="https://jsss.copernicus.org/articles/5/261/2016/jsss-5-261-2016-g01.pdf"/> (60 s
recording time per laser).</p></caption>
          <?xmltex \igopts{width=236.157874pt}?><graphic xlink:href="https://jsss.copernicus.org/articles/5/261/2016/jsss-5-261-2016-f13.png"/>

        </fig>

      <?xmltex \floatpos{t}?><fig id="Ch1.F14" specific-use="star"><caption><p>Piece of sugar in water. The dissolving process is tracked
using quasi real-time Raman imaging: first frame (left, 0 s), 20th frame
(center, 390 s), and 40th frame (right, 780 s). The pseudo colors illustrate
the intensity at 850 cm<inline-formula><mml:math display="inline"><mml:msup><mml:mi/><mml:mrow><mml:mo>-</mml:mo><mml:mn mathvariant="normal">1</mml:mn></mml:mrow></mml:msup></mml:math></inline-formula>. Below the corresponding Raman spectra at the
spatial position <inline-formula><mml:math display="inline"><mml:mrow><mml:mo>(</mml:mo><mml:mi>X</mml:mi><mml:mo>/</mml:mo><mml:mi>Y</mml:mi><mml:mo>)</mml:mo><mml:mo>=</mml:mo><mml:mo>(</mml:mo><mml:mn mathvariant="normal">7</mml:mn><mml:mo>/</mml:mo><mml:mn mathvariant="normal">8</mml:mn><mml:mo>)</mml:mo></mml:mrow></mml:math></inline-formula> are shown.</p></caption>
          <?xmltex \igopts{width=455.244094pt}?><graphic xlink:href="https://jsss.copernicus.org/articles/5/261/2016/jsss-5-261-2016-f14.png"/>

        </fig>

</sec>
<sec id="Ch1.S3.SS4">
  <title>Cut surface of a porcine ear</title>
      <p>From a fresh porcine ear, i.e., the ear was not scalded after slaughtering,
the upper part was cut with a scalpel and placed face down at the calcium
fluoride bottom of a specimen vessel (Fig. 11). The cut face was centered and
focused. The Raman spectra were recorded at 784.5 and 785.5 nm excitation
wavelengths for 60 s each. In the middle of the cross section there is
obviously chondral tissue. The camera image (Fig. 12) shows it as a 3 mm
broad and light vertical stripe that is slightly tilted (region between the
dotted lines). Around the <inline-formula><mml:math display="inline"><mml:mrow><mml:mo>(</mml:mo><mml:mi>X</mml:mi><mml:mo>/</mml:mo><mml:mi>Y</mml:mi><mml:mo>)</mml:mo></mml:mrow></mml:math></inline-formula> positions <inline-formula><mml:math display="inline"><mml:mrow><mml:mo>(</mml:mo><mml:mn mathvariant="normal">7</mml:mn><mml:mo>/</mml:mo><mml:mn>10</mml:mn><mml:mo>)</mml:mo></mml:mrow></mml:math></inline-formula> and <inline-formula><mml:math display="inline"><mml:mrow><mml:mo>(</mml:mo><mml:mn mathvariant="normal">9</mml:mn><mml:mo>/</mml:mo><mml:mn mathvariant="normal">7</mml:mn><mml:mo>)</mml:mo></mml:mrow></mml:math></inline-formula> there are
reddish spots and at <inline-formula><mml:math display="inline"><mml:mrow><mml:mo>(</mml:mo><mml:mn>16</mml:mn><mml:mo>/</mml:mo><mml:mn mathvariant="normal">3</mml:mn><mml:mo>)</mml:mo></mml:mrow></mml:math></inline-formula> there is a dark spot (marked by the dotted
circle in the camera image). The position of the chondral tissue as well as
those of the reddish and dark spots were retrieved within the Raman image
that corresponds to the intensities at 1303 cm<inline-formula><mml:math display="inline"><mml:msup><mml:mi/><mml:mrow><mml:mo>-</mml:mo><mml:mn mathvariant="normal">1</mml:mn></mml:mrow></mml:msup></mml:math></inline-formula> (lipid band).</p>
</sec>
<sec id="Ch1.S3.SS5">
  <title>Porcine skin</title>
      <p>A piece of porcine skin is placed with the surface down on the 1 mm calcium
fluoride slide and focused. Porcine skin is morphologically similar to human
skin (Darvin et al., 2014; Tfaili et al., 2012), and shall be used as a model
system to optimize the spectrograph with regard to future human skin cancer
diagnostics. To avoid drying of the skin, some water was supplied on the
slide and the specimen vessel closed with a lid. The Raman spectra were
recorded at 784.5 and 785.5 nm excitation wavelengths for 60 s each. The
results are shown in Fig. 13. The Raman image was generated by using the
Raman intensity at 1455 cm<inline-formula><mml:math display="inline"><mml:msup><mml:mi/><mml:mrow><mml:mo>-</mml:mo><mml:mn mathvariant="normal">1</mml:mn></mml:mrow></mml:msup></mml:math></inline-formula> (CH<inline-formula><mml:math display="inline"><mml:msub><mml:mi/><mml:mn mathvariant="normal">2</mml:mn></mml:msub></mml:math></inline-formula> bending vibrations of proteins).
The structures that are visible in the camera image also find themselves in
the Raman image in detail. The pigmented spot in the upper right corner as
well as the obviously somewhat dirty hair that passes diagonally the upper
right quadrant of the image field are very visible. The weak pigmented spots
at <inline-formula><mml:math display="inline"><mml:mrow><mml:mo>(</mml:mo><mml:mn mathvariant="normal">3</mml:mn><mml:mo>/</mml:mo><mml:mn>15</mml:mn><mml:mo>)</mml:mo></mml:mrow></mml:math></inline-formula> and <inline-formula><mml:math display="inline"><mml:mrow><mml:mo>(</mml:mo><mml:mn>13</mml:mn><mml:mo>/</mml:mo><mml:mn mathvariant="normal">5</mml:mn><mml:mo>)</mml:mo></mml:mrow></mml:math></inline-formula> are also identifiable in the Raman image. The
obtained Raman spectra are similar, with Raman spectra of porcine skin
obtained using another Raman microscope (Zhu et al., 2015).<?xmltex \hack{\newpage}?></p>
</sec>
<sec id="Ch1.S3.SS6">
  <title>Piece of brown sugar</title>
      <p>With regard to future real-time Raman imaging of samples that change over
time, a software script was applied that allows automatic recording of Raman
spectra sequences. As a model system, the dissolution of a piece of brown
sugar in water was chosen. The sugar was put into a water-filled dish at the
top of AO2 and the Raman measurements were started. For each frame, the
recording time was 10 s, followed by 10 s read-out time. Thus, the
effective temporal resolution was 20 s. To track the dissolving process of
the sugar in quasi real time, a total number of 60 frames was recorded.
Figure 14 shows the situation at 0, 390 and 780 s. At the top Raman and
camera images, at the bottom the corresponding Raman spectra at the spatial
position <inline-formula><mml:math display="inline"><mml:mrow><mml:mo>(</mml:mo><mml:mi>X</mml:mi><mml:mo>/</mml:mo><mml:mi>Y</mml:mi><mml:mo>)</mml:mo><mml:mo>=</mml:mo><mml:mo>(</mml:mo><mml:mn mathvariant="normal">7</mml:mn><mml:mo>/</mml:mo><mml:mn mathvariant="normal">8</mml:mn><mml:mo>)</mml:mo></mml:mrow></mml:math></inline-formula> are displayed. The rolling circle method (Brandt et
al., 2006) was applied to remove background. The spectra reveal that the
sugar piece consists of sucrose (Brizuela et al., 2012). The pseudo color
plots illustrate the intensity of the strongest Raman signal at
850 cm<inline-formula><mml:math display="inline"><mml:msup><mml:mi/><mml:mrow><mml:mo>-</mml:mo><mml:mn mathvariant="normal">1</mml:mn></mml:mrow></mml:msup></mml:math></inline-formula> (CH<inline-formula><mml:math display="inline"><mml:msub><mml:mi/><mml:mn mathvariant="normal">2</mml:mn></mml:msub></mml:math></inline-formula> twisting). The Raman images perfectly match the
shape and position of the sugar piece shown at the camera images.</p>
      <p>The bottleneck with regard to the temporal resolution is the long read-out
time of the CCD chip. Unfortunately the available camera system always
requires an entire read-out of the CCD chip. However, a new camera system
that allows one to define custom read-out regions and thus shorter down times
is currently under construction.</p>
</sec>
</sec>
<sec id="Ch1.S4" sec-type="conclusions">
  <title>Conclusions</title>
      <p>The setup described in this paper allows the recording of spatially resolved
large-scale (up to 100 mm<inline-formula><mml:math display="inline"><mml:mrow><mml:msup><mml:mi/><mml:mn mathvariant="normal">2</mml:mn></mml:msup><mml:mo>)</mml:mo></mml:mrow></mml:math></inline-formula> Raman images without a scanning procedure.
The exposure times to receive the fill spectral and spatial information were
120 s (cut face of porcine ear and porcine skin), 20 s (silicon and EUR 10
bill), 10 s (sugar) and 2 s (Aspirin Effect), which corresponds to 300, 50,
25, and 5 ms, respectively, per pixel. The read-out time of the CCD chip was
10 s. To receive background-free Raman spectra via SERDS, two successive
exposures and read-outs are required. The pseudo color Raman images display
the intensities of selected bands of the Raman spectra at the corresponding
spatial positions. Even with this quite ordinary analysis method,
high-contrast Raman images are gained that correspond well to the structures
shown by the camera. The results show great promise for being able to
identify cancerous tissue in a next step by using differential analysis.</p>
</sec>

      
      </body>
    <back><notes notes-type="authorcontribution">

      <p>E. Schmälzlin, B. Moralejo, and M. M. Roth designed and realized the
optical setup. E. Schmälzlin, B. Moralejo, D. Bodenmüller, M. E.
Darvin, and G. Thiede designed and carried out the experiments. E.
Schmälzlin, B. Moralejo, and M. E. Darvin analyzed data. E.
Schmälzlin and B. Moralejo prepared the manuscript. All authors have
revised the study critically and approved submission.</p>
  </notes><ack><title>Acknowledgements</title><p>The authors acknowledge support from the German Federal Ministry of Education
and Research (BMBF) VIP program, grant no. 03V0843, and the Zwanzig20
program, grant no. 03ZZ0423. We thank Johannes Schleusener, Ingo Gersonde,
and Jürgen Helfmann from Laser- und Medizin-Technologie GmbH, Berlin, 14195, Germany,
for their assistance by the development of the MLA-based optic unit.
<?xmltex \hack{\newline}?><?xmltex \hack{\newline}?> Edited by: G. Gauglitz<?xmltex \hack{\newline}?>
Reviewed by: two anonymous referees</p></ack><ref-list>
    <title>References</title>

      <ref id="bib1.bib1"><label>1</label><mixed-citation>Allington-Smith, J.: Basic principles of integral field spec<?xmltex \hack{\-}?>troscopy, New
Astron. Rev., 50, 244–251, <ext-link xlink:href="http://dx.doi.org/10.1016/j.newar.2006.02.024" ext-link-type="DOI">10.1016/j.newar.2006.02.024</ext-link>, 2006.</mixed-citation></ref>
      <ref id="bib1.bib2"><label>2</label><mixed-citation>Bacon, R., Bauer, S.-M., Bower, R., Cabrit, S., Cappellari, M., Carollo, M.,
Combes, F., Davies, R. L., Delabre, B., Dekker, H., Devriendt, J., Djidel,
S., Duchateau, M., Dubois, J.-P., Emsellem, E., Ferruit, P., Franx, M.,
Gilmore, G. F., Guiderdoni, B., Henault, F., Hubin, N., Jungwiert, B., Kelz,
A., Le Louarn, M., Lewis, I. J., Lizon, J.-L., McDermid, R., Morris, S. L.,
Laux, U., Le Fèvre, O., Lantz, B., Lilly, S., Lynn, J., Pasquini, L.,
Pecontal, A., Pinet, P., Popovic, D., Quirrenbach, A., Reiss, R., Roth, M.
M., Steinmetz, M., Stuik, R., Wisotzki, L., and de Zeeuw, P. T.: The
second-generation VLT instrument MUSE: science drivers and instrument
design, in: Proc. SPIE 5492, Ground-based Instrumentation for Astronomy,
Glasgow, United Kingdom, 21–25 June 2004, 1145, <ext-link xlink:href="http://dx.doi.org/10.1117/12.549009" ext-link-type="DOI">10.1117/12.549009</ext-link>,
2004.</mixed-citation></ref>
      <ref id="bib1.bib3"><label>3</label><mixed-citation>Brandt, N. N., Brovko, O. O., Chikishev, A. Y., and Paraschuk, O. D.:
Optimization of the Rolling-Circle Filter for Raman Background Subtraction,
Appl. Spectrosc., 60, 288–293, <ext-link xlink:href="http://dx.doi.org/10.1366/000370206776342553" ext-link-type="DOI">10.1366/000370206776342553</ext-link>, 2006.</mixed-citation></ref>
      <ref id="bib1.bib4"><label>4</label><mixed-citation>Brizuela, A. B., Bichara, L. C., Romano, E., Yurquina, A., Locatelli, S.,
and Brandán, S. A.: A complete characterization of the vibrational
spectra of sucrose, Carbohyd. Res., 361, 212–218, <ext-link xlink:href="http://dx.doi.org/10.1016/j.carres.2012.07.009" ext-link-type="DOI">10.1016/j.carres.2012.07.009</ext-link>, 2012.</mixed-citation></ref>
      <ref id="bib1.bib5"><label>5</label><mixed-citation>Brückner, M., Becker, K., Popp, J., and Frosch, T.: Fiber array based
hyperspectral Raman imaging for chemical selective analysis of
malaria-infected red blood cells, Anal. Chim. Acta, 894, 76–84,
<ext-link xlink:href="http://dx.doi.org/10.1016/j.aca.2015.08.025" ext-link-type="DOI">10.1016/j.aca.2015.08.025</ext-link>, 2015.</mixed-citation></ref>
      <ref id="bib1.bib6"><label>6</label><mixed-citation>Darvin, M. E., Richter, H., Zhu, Y. J., Meinke, M. C., Knorr, F., Gonchukov,
S. A., Koenig, K., and Lademann, J.: Comparison of in vivo and ex vivo laser
scanning microscopy and multiphoton tomography application for human and
porcine skin imaging, Quantum Electron., 44, 646–651,
<ext-link xlink:href="http://dx.doi.org/10.1070/QE2014v044n07ABEH015488" ext-link-type="DOI">10.1070/QE2014v044n07ABEH015488</ext-link>, 2014.</mixed-citation></ref>
      <ref id="bib1.bib7"><label>7</label><mixed-citation>Hagen, N., Kester, R. T., Gao, L., and Tkaczyk, T. S.: Snapshot advantage: a
review of the light collection improvement for parallel high-dimensional
measurement systems, Opt. Eng., 51, 111702, <ext-link xlink:href="http://dx.doi.org/10.1117/1.OE.51.11.111702" ext-link-type="DOI">10.1117/1.OE.51.11.111702</ext-link>,
2012.</mixed-citation></ref>
      <ref id="bib1.bib8"><label>8</label><mixed-citation>Li, Q., He, X., Wang, Y., Liu, H., Xu, D., and Guo, F.: Review of spectral
imaging technology in biomedical engineering: achievements and challenges,
J. Biomed. Opt., 18, 100901, <ext-link xlink:href="http://dx.doi.org/10.1117/1.JBO.18.10.100901" ext-link-type="DOI">10.1117/1.JBO.18.10.100901</ext-link>, 2013.</mixed-citation></ref>
      <ref id="bib1.bib9"><label>9</label><mixed-citation>Ma, J. and Ben-Amotz, D.: Rapid Micro-Raman Imaging Using Fiber-Bundle Image
Compression, Appl. Spectrosc., 51, 1845–1848, <ext-link xlink:href="http://dx.doi.org/10.1366/0003702971939668" ext-link-type="DOI">10.1366/0003702971939668</ext-link>, 1997.</mixed-citation></ref>
      <ref id="bib1.bib10"><label>10</label><mixed-citation>Okuno, M. and Hamaguchi, H.: Multifocus confocal Raman microspectroscopy for
fast multimode vibrational imaging of living cells, Opt. Lett., 35,
4096–4098, <ext-link xlink:href="http://dx.doi.org/10.1364/OL.35.004096" ext-link-type="DOI">10.1364/OL.35.004096</ext-link>, 2010.</mixed-citation></ref>
      <ref id="bib1.bib11"><label>11</label><mixed-citation>Opilik, L., Schmid, T., and Zenobi, R.: Modern Raman Imaging: Vibrational
Spectroscopy on the Micrometer and Nanometer Scales, Annu. Rev. Anal. Chem.,
6, 379–398, <ext-link xlink:href="http://dx.doi.org/10.1146/annurev-anchem-062012-092646" ext-link-type="DOI">10.1146/annurev-anchem-062012-092646</ext-link>, 2013.</mixed-citation></ref>
      <ref id="bib1.bib12"><label>12</label><mixed-citation>Parker Jr., J. H., Feldman, D. W., and Ashkin, M.; Raman Scattering by
Silicon and Germanium, Phys. Rev., 155, 712–714, <ext-link xlink:href="http://dx.doi.org/10.1103/PhysRev.155.712" ext-link-type="DOI">10.1103/PhysRev.155.712</ext-link>, 1967.</mixed-citation></ref>
      <ref id="bib1.bib13"><label>13</label><mixed-citation>Reble, C., Gersonde, I., Dressler, C., Helfmann, J., Kühn, W., and
Eichler, H. J.: Evaluation of Raman spectroscopic macro raster scans of
native cervical cone biopsies using histopathological mapping, J. Biomed.
Opt., 19, 027007, <ext-link xlink:href="http://dx.doi.org/10.1117/1.JBO.19.2.027007" ext-link-type="DOI">10.1117/1.JBO.19.2.027007</ext-link>, 2014.</mixed-citation></ref>
      <ref id="bib1.bib14"><label>14</label><mixed-citation>
Roth, M.: Introductory review and technical approaches, in: 3D Spectroscopy
in Astronomy, edited by: Mediavilla, E., Arribas, S., Roth, M., Cepa-Norgué, and
Sánchez, F., Cambridge University Press, New York, USA, 1–39,
2010.</mixed-citation></ref>
      <ref id="bib1.bib15"><label>15</label><mixed-citation>Roth, M. M., Zenichowski, K., Tarcea, N., Popp, J., Adelhelm, S., Stolz, M.,
Kelz, A., Sandin, C., Bauer, S.-M., Fechner, T., Popow, E., Roth, B., Singh,
P., Srivastava, M., and Wolter, D.: The ERA2 facility: towards application of a
fibre-based astronomical spectrograph for imaging spectroscopy in life
sciences, in: Proc. SPIE 8450, Modern Technologies in Space- and
Ground-based Telescopes and Instrumentation II, Amsterdam, the Netherlands,
1 July 2012, 84501T, <ext-link xlink:href="http://dx.doi.org/10.1117/12.925340" ext-link-type="DOI">10.1117/12.925340</ext-link>, 2012</mixed-citation></ref>
      <ref id="bib1.bib16"><label>16</label><mixed-citation>Sandin, C., Becker, T., Roth, M. M., Gerssen, J., Monreal-Ibero, A.,
Böhm, P., and Weilbacher, P.: p3d: a general data-reduction tool for
fiber-fed integral-field spectrographs, Astron. Astrophys., 515, A35,
<ext-link xlink:href="http://dx.doi.org/10.1051/0004-6361/201014022" ext-link-type="DOI">10.1051/0004-6361/201014022</ext-link>, 2010.</mixed-citation></ref>
      <ref id="bib1.bib17"><label>17</label><mixed-citation>Schmälzlin, E., Moralejo, B., Rutowska, M., Monreal-Ibero, A., Sandin,
C., Tarcea, N., Popp J., and Roth M. M.: Raman Imaging with a Fiber-Coupled
Multichannel Spectrograph, Sensors, 14, 21968–21980, <ext-link xlink:href="http://dx.doi.org/10.3390/s141121968" ext-link-type="DOI">10.3390/s141121968</ext-link>, 2014.</mixed-citation></ref>
      <ref id="bib1.bib18"><label>18</label><mixed-citation>
Schmälzlin, E., Stolz, M., Moralejo, B., Adelhelm, S., and Roth, M. M.:
Imaging multiplex Raman spectroscopy with no need for sequential scanning,
Laser. Photonics Rev., 62–65, 2015.</mixed-citation></ref>
      <ref id="bib1.bib19"><label>19</label><mixed-citation>
Schmidt, H., Kaiser, P., and Maiwald, M.: European patent EP 2 480 868 B1,
Patent bulletin 2014/02, 2014.</mixed-citation></ref>
      <ref id="bib1.bib20"><label>20</label><mixed-citation>Sowoidnich, K. and Kronfeldt, H.-D.: Fluorescence Rejection by Shifted
Excitation Raman Difference Spectroscopy at Multiple Wavelengths for the
Investigation of Biological Samples, ISRN Spectroscopy 2012, 256326,
<ext-link xlink:href="http://dx.doi.org/10.5402/2012/256326" ext-link-type="DOI">10.5402/2012/256326</ext-link>, 2012.</mixed-citation></ref>
      <ref id="bib1.bib21"><label>21</label><mixed-citation>Stewart S., Priore, R. J., Nelson, M. P., and Treado, P. J.: Raman Imaging,
Annu. Rev. Anal. Chem. 5, 337–260, <ext-link xlink:href="http://dx.doi.org/10.1146/annurev-anchem-062011-143152" ext-link-type="DOI">10.1146/annurev-anchem-062011-143152</ext-link>, 2012.</mixed-citation></ref>
      <ref id="bib1.bib22"><label>22</label><mixed-citation>Tfaili, S., Gobinet, C., Josse, G., Angiboust, J.-F., Manfait, M., and Piot,
O.: Confocal Raman microspectroscopy for skin characterization: a
comparative study between human skin and pig skin, Analyst, 137, 3673–3682,
<ext-link xlink:href="http://dx.doi.org/10.1039/c2an16292j" ext-link-type="DOI">10.1039/c2an16292j</ext-link>, 2012.
</mixed-citation></ref><?xmltex \hack{\newpage}?>
      <ref id="bib1.bib23"><label>23</label><mixed-citation>Vanderriest, C.: Fiber-Optics Dissector for Spectroscopy of Nebulosities
around Quasars and similar Objects, PASP, 92, 858–862, <ext-link xlink:href="http://dx.doi.org/10.1086/130764" ext-link-type="DOI">10.1086/130764</ext-link>,
1980.</mixed-citation></ref>
      <ref id="bib1.bib24"><label>24</label><mixed-citation>Wei, D., Chen, S., and Liu, Q.: Review of Fluorescence Suppression
Techniques in Raman Spectroscopy, Appl. Spectrosc. Rev., 50, 387–406,
<ext-link xlink:href="http://dx.doi.org/10.1080/05704928.2014.999936" ext-link-type="DOI">10.1080/05704928.2014.999936</ext-link>, 2015.</mixed-citation></ref>
      <ref id="bib1.bib25"><label>25</label><mixed-citation>Zhao, J., Lui, H., Kalia, S., and Zeng, H.: Real-time Raman spectroscopy for
automatic in vivo skin cancer detection: an independent validation, Anal.
Bioanal. Chem., 407, 8373–8379, <ext-link xlink:href="http://dx.doi.org/10.1007/s00216-015-8914-9" ext-link-type="DOI">10.1007/s00216-015-8914-9</ext-link>, 2015.</mixed-citation></ref>
      <ref id="bib1.bib26"><label>26</label><mixed-citation>Zhu, Y. J., Choe, C.-S., Ahlberg, S., Lademann, J., and Darvin, M. E.:
Penetration of silver nanoparticles into porcine skin ex vivo using
fluorescence lifetime imaging microscopy, Raman microscopy and surface
enhanced Raman microscopy, J. Biomed. Opt., 20, 051006,
<ext-link xlink:href="http://dx.doi.org/10.1117/1.JBO.20.5.051006" ext-link-type="DOI">10.1117/1.JBO.20.5.051006</ext-link>, 2015.</mixed-citation></ref>

  </ref-list><app-group content-type="float"><app><title/>

    </app></app-group></back>
    <!--<article-title-html>Ultrafast imaging Raman spectroscopy of large-area samples without stepwise
scanning</article-title-html>
<abstract-html><p class="p">Step-by-step, time-consuming scanning of the sample is still the
state-of-the-art in imaging Raman spectroscopy. Even for a few 100 image
points the measurement time may add up to minutes or hours. A radical
decrease in measurement time can be achieved by applying multiplex
spectrographs coupled to imaging fiber bundles that are successfully used in
astronomy. For optimal use of the scarce and expensive observation time at
astronomical observatories, special high-performance spectrograph systems
were developed. They are designed for recording thousands of spatially
resolved spectra of a two-dimensional image field within one single exposure.
Transferring this technology to imaging Raman spectroscopy allows a
considerably faster acquisition of chemical maps. Currently, an imaging field
of up to 1 cm<sup>2</sup> can be investigated. For porcine skin the required
measurement time is less than 1 min. For this reason, this technique is of
particular interest for medical diagnostics, e.g., the identification of
potentially cancerous abnormalities of skin tissue.</p></abstract-html>
<ref-html id="bib1.bib1"><label>1</label><mixed-citation>
Allington-Smith, J.: Basic principles of integral field spectroscopy, New
Astron. Rev., 50, 244–251, <a href="http://dx.doi.org/10.1016/j.newar.2006.02.024" target="_blank">doi:10.1016/j.newar.2006.02.024</a>, 2006.
</mixed-citation></ref-html>
<ref-html id="bib1.bib2"><label>2</label><mixed-citation>
Bacon, R., Bauer, S.-M., Bower, R., Cabrit, S., Cappellari, M., Carollo, M.,
Combes, F., Davies, R. L., Delabre, B., Dekker, H., Devriendt, J., Djidel,
S., Duchateau, M., Dubois, J.-P., Emsellem, E., Ferruit, P., Franx, M.,
Gilmore, G. F., Guiderdoni, B., Henault, F., Hubin, N., Jungwiert, B., Kelz,
A., Le Louarn, M., Lewis, I. J., Lizon, J.-L., McDermid, R., Morris, S. L.,
Laux, U., Le Fèvre, O., Lantz, B., Lilly, S., Lynn, J., Pasquini, L.,
Pecontal, A., Pinet, P., Popovic, D., Quirrenbach, A., Reiss, R., Roth, M.
M., Steinmetz, M., Stuik, R., Wisotzki, L., and de Zeeuw, P. T.: The
second-generation VLT instrument MUSE: science drivers and instrument
design, in: Proc. SPIE 5492, Ground-based Instrumentation for Astronomy,
Glasgow, United Kingdom, 21–25 June 2004, 1145, <a href="http://dx.doi.org/10.1117/12.549009" target="_blank">doi:10.1117/12.549009</a>,
2004.
</mixed-citation></ref-html>
<ref-html id="bib1.bib3"><label>3</label><mixed-citation>
Brandt, N. N., Brovko, O. O., Chikishev, A. Y., and Paraschuk, O. D.:
Optimization of the Rolling-Circle Filter for Raman Background Subtraction,
Appl. Spectrosc., 60, 288–293, <a href="http://dx.doi.org/10.1366/000370206776342553" target="_blank">doi:10.1366/000370206776342553</a>, 2006.
</mixed-citation></ref-html>
<ref-html id="bib1.bib4"><label>4</label><mixed-citation>
Brizuela, A. B., Bichara, L. C., Romano, E., Yurquina, A., Locatelli, S.,
and Brandán, S. A.: A complete characterization of the vibrational
spectra of sucrose, Carbohyd. Res., 361, 212–218, <a href="http://dx.doi.org/10.1016/j.carres.2012.07.009" target="_blank">doi:10.1016/j.carres.2012.07.009</a>, 2012.
</mixed-citation></ref-html>
<ref-html id="bib1.bib5"><label>5</label><mixed-citation>
Brückner, M., Becker, K., Popp, J., and Frosch, T.: Fiber array based
hyperspectral Raman imaging for chemical selective analysis of
malaria-infected red blood cells, Anal. Chim. Acta, 894, 76–84,
<a href="http://dx.doi.org/10.1016/j.aca.2015.08.025" target="_blank">doi:10.1016/j.aca.2015.08.025</a>, 2015.
</mixed-citation></ref-html>
<ref-html id="bib1.bib6"><label>6</label><mixed-citation>
Darvin, M. E., Richter, H., Zhu, Y. J., Meinke, M. C., Knorr, F., Gonchukov,
S. A., Koenig, K., and Lademann, J.: Comparison of in vivo and ex vivo laser
scanning microscopy and multiphoton tomography application for human and
porcine skin imaging, Quantum Electron., 44, 646–651,
<a href="http://dx.doi.org/10.1070/QE2014v044n07ABEH015488" target="_blank">doi:10.1070/QE2014v044n07ABEH015488</a>, 2014.
</mixed-citation></ref-html>
<ref-html id="bib1.bib7"><label>7</label><mixed-citation>
Hagen, N., Kester, R. T., Gao, L., and Tkaczyk, T. S.: Snapshot advantage: a
review of the light collection improvement for parallel high-dimensional
measurement systems, Opt. Eng., 51, 111702, <a href="http://dx.doi.org/10.1117/1.OE.51.11.111702" target="_blank">doi:10.1117/1.OE.51.11.111702</a>,
2012.
</mixed-citation></ref-html>
<ref-html id="bib1.bib8"><label>8</label><mixed-citation>
Li, Q., He, X., Wang, Y., Liu, H., Xu, D., and Guo, F.: Review of spectral
imaging technology in biomedical engineering: achievements and challenges,
J. Biomed. Opt., 18, 100901, <a href="http://dx.doi.org/10.1117/1.JBO.18.10.100901" target="_blank">doi:10.1117/1.JBO.18.10.100901</a>, 2013.
</mixed-citation></ref-html>
<ref-html id="bib1.bib9"><label>9</label><mixed-citation>
Ma, J. and Ben-Amotz, D.: Rapid Micro-Raman Imaging Using Fiber-Bundle Image
Compression, Appl. Spectrosc., 51, 1845–1848, <a href="http://dx.doi.org/10.1366/0003702971939668" target="_blank">doi:10.1366/0003702971939668</a>, 1997.
</mixed-citation></ref-html>
<ref-html id="bib1.bib10"><label>10</label><mixed-citation>
Okuno, M. and Hamaguchi, H.: Multifocus confocal Raman microspectroscopy for
fast multimode vibrational imaging of living cells, Opt. Lett., 35,
4096–4098, <a href="http://dx.doi.org/10.1364/OL.35.004096" target="_blank">doi:10.1364/OL.35.004096</a>, 2010.
</mixed-citation></ref-html>
<ref-html id="bib1.bib11"><label>11</label><mixed-citation>
Opilik, L., Schmid, T., and Zenobi, R.: Modern Raman Imaging: Vibrational
Spectroscopy on the Micrometer and Nanometer Scales, Annu. Rev. Anal. Chem.,
6, 379–398, <a href="http://dx.doi.org/10.1146/annurev-anchem-062012-092646" target="_blank">doi:10.1146/annurev-anchem-062012-092646</a>, 2013.
</mixed-citation></ref-html>
<ref-html id="bib1.bib12"><label>12</label><mixed-citation>
Parker Jr., J. H., Feldman, D. W., and Ashkin, M.; Raman Scattering by
Silicon and Germanium, Phys. Rev., 155, 712–714, <a href="http://dx.doi.org/10.1103/PhysRev.155.712" target="_blank">doi:10.1103/PhysRev.155.712</a>, 1967.
</mixed-citation></ref-html>
<ref-html id="bib1.bib13"><label>13</label><mixed-citation>
Reble, C., Gersonde, I., Dressler, C., Helfmann, J., Kühn, W., and
Eichler, H. J.: Evaluation of Raman spectroscopic macro raster scans of
native cervical cone biopsies using histopathological mapping, J. Biomed.
Opt., 19, 027007, <a href="http://dx.doi.org/10.1117/1.JBO.19.2.027007" target="_blank">doi:10.1117/1.JBO.19.2.027007</a>, 2014.
</mixed-citation></ref-html>
<ref-html id="bib1.bib14"><label>14</label><mixed-citation>
Roth, M.: Introductory review and technical approaches, in: 3D Spectroscopy
in Astronomy, edited by: Mediavilla, E., Arribas, S., Roth, M., Cepa-Norgué, and
Sánchez, F., Cambridge University Press, New York, USA, 1–39,
2010.
</mixed-citation></ref-html>
<ref-html id="bib1.bib15"><label>15</label><mixed-citation>
Roth, M. M., Zenichowski, K., Tarcea, N., Popp, J., Adelhelm, S., Stolz, M.,
Kelz, A., Sandin, C., Bauer, S.-M., Fechner, T., Popow, E., Roth, B., Singh,
P., Srivastava, M., and Wolter, D.: The ERA2 facility: towards application of a
fibre-based astronomical spectrograph for imaging spectroscopy in life
sciences, in: Proc. SPIE 8450, Modern Technologies in Space- and
Ground-based Telescopes and Instrumentation II, Amsterdam, the Netherlands,
1 July 2012, 84501T, <a href="http://dx.doi.org/10.1117/12.925340" target="_blank">doi:10.1117/12.925340</a>, 2012
</mixed-citation></ref-html>
<ref-html id="bib1.bib16"><label>16</label><mixed-citation>
Sandin, C., Becker, T., Roth, M. M., Gerssen, J., Monreal-Ibero, A.,
Böhm, P., and Weilbacher, P.: p3d: a general data-reduction tool for
fiber-fed integral-field spectrographs, Astron. Astrophys., 515, A35,
<a href="http://dx.doi.org/10.1051/0004-6361/201014022" target="_blank">doi:10.1051/0004-6361/201014022</a>, 2010.
</mixed-citation></ref-html>
<ref-html id="bib1.bib17"><label>17</label><mixed-citation>
Schmälzlin, E., Moralejo, B., Rutowska, M., Monreal-Ibero, A., Sandin,
C., Tarcea, N., Popp J., and Roth M. M.: Raman Imaging with a Fiber-Coupled
Multichannel Spectrograph, Sensors, 14, 21968–21980, <a href="http://dx.doi.org/10.3390/s141121968" target="_blank">doi:10.3390/s141121968</a>, 2014.
</mixed-citation></ref-html>
<ref-html id="bib1.bib18"><label>18</label><mixed-citation>
Schmälzlin, E., Stolz, M., Moralejo, B., Adelhelm, S., and Roth, M. M.:
Imaging multiplex Raman spectroscopy with no need for sequential scanning,
Laser. Photonics Rev., 62–65, 2015.
</mixed-citation></ref-html>
<ref-html id="bib1.bib19"><label>19</label><mixed-citation>
Schmidt, H., Kaiser, P., and Maiwald, M.: European patent EP 2 480 868 B1,
Patent bulletin 2014/02, 2014.
</mixed-citation></ref-html>
<ref-html id="bib1.bib20"><label>20</label><mixed-citation>
Sowoidnich, K. and Kronfeldt, H.-D.: Fluorescence Rejection by Shifted
Excitation Raman Difference Spectroscopy at Multiple Wavelengths for the
Investigation of Biological Samples, ISRN Spectroscopy 2012, 256326,
<a href="http://dx.doi.org/10.5402/2012/256326" target="_blank">doi:10.5402/2012/256326</a>, 2012.
</mixed-citation></ref-html>
<ref-html id="bib1.bib21"><label>21</label><mixed-citation>
Stewart S., Priore, R. J., Nelson, M. P., and Treado, P. J.: Raman Imaging,
Annu. Rev. Anal. Chem. 5, 337–260, <a href="http://dx.doi.org/10.1146/annurev-anchem-062011-143152" target="_blank">doi:10.1146/annurev-anchem-062011-143152</a>, 2012.
</mixed-citation></ref-html>
<ref-html id="bib1.bib22"><label>22</label><mixed-citation>
Tfaili, S., Gobinet, C., Josse, G., Angiboust, J.-F., Manfait, M., and Piot,
O.: Confocal Raman microspectroscopy for skin characterization: a
comparative study between human skin and pig skin, Analyst, 137, 3673–3682,
<a href="http://dx.doi.org/10.1039/c2an16292j" target="_blank">doi:10.1039/c2an16292j</a>, 2012.

</mixed-citation></ref-html>
<ref-html id="bib1.bib23"><label>23</label><mixed-citation>
Vanderriest, C.: Fiber-Optics Dissector for Spectroscopy of Nebulosities
around Quasars and similar Objects, PASP, 92, 858–862, <a href="http://dx.doi.org/10.1086/130764" target="_blank">doi:10.1086/130764</a>,
1980.
</mixed-citation></ref-html>
<ref-html id="bib1.bib24"><label>24</label><mixed-citation>
Wei, D., Chen, S., and Liu, Q.: Review of Fluorescence Suppression
Techniques in Raman Spectroscopy, Appl. Spectrosc. Rev., 50, 387–406,
<a href="http://dx.doi.org/10.1080/05704928.2014.999936" target="_blank">doi:10.1080/05704928.2014.999936</a>, 2015.
</mixed-citation></ref-html>
<ref-html id="bib1.bib25"><label>25</label><mixed-citation>
Zhao, J., Lui, H., Kalia, S., and Zeng, H.: Real-time Raman spectroscopy for
automatic in vivo skin cancer detection: an independent validation, Anal.
Bioanal. Chem., 407, 8373–8379, <a href="http://dx.doi.org/10.1007/s00216-015-8914-9" target="_blank">doi:10.1007/s00216-015-8914-9</a>, 2015.
</mixed-citation></ref-html>
<ref-html id="bib1.bib26"><label>26</label><mixed-citation>
Zhu, Y. J., Choe, C.-S., Ahlberg, S., Lademann, J., and Darvin, M. E.:
Penetration of silver nanoparticles into porcine skin ex vivo using
fluorescence lifetime imaging microscopy, Raman microscopy and surface
enhanced Raman microscopy, J. Biomed. Opt., 20, 051006,
<a href="http://dx.doi.org/10.1117/1.JBO.20.5.051006" target="_blank">doi:10.1117/1.JBO.20.5.051006</a>, 2015.
</mixed-citation></ref-html>--></article>