Articles | Volume 10, issue 1
https://doi.org/10.5194/jsss-10-109-2021
https://doi.org/10.5194/jsss-10-109-2021
Regular research article
 | 
23 Apr 2021
Regular research article |  | 23 Apr 2021

Absolute calibration of the spectral responsivity of thermal detectors in the near-infrared (NIR) and mid-infrared (MIR) regions by using blackbody radiation

Tobias Pohl, Peter Meindl, Lutz Werner, Uwe Johannsen, Dieter Taubert, Christian Monte, and Jörg Hollandt

Viewed

Total article views: 1,437 (including HTML, PDF, and XML)
HTML PDF XML Total BibTeX EndNote
787 613 37 1,437 23 25
  • HTML: 787
  • PDF: 613
  • XML: 37
  • Total: 1,437
  • BibTeX: 23
  • EndNote: 25
Views and downloads (calculated since 23 Apr 2021)
Cumulative views and downloads (calculated since 23 Apr 2021)

Viewed (geographical distribution)

Total article views: 1,331 (including HTML, PDF, and XML) Thereof 1,331 with geography defined and 0 with unknown origin.
Country # Views %
  • 1
1
 
 
 
 

Cited

Latest update: 23 Nov 2024
Download
Short summary
The Physikalisch-Technische Bundesanstalt has set up a primary method for the calibration of the spectral responsivity of thermal detectors in the near- and mid-infrared spectral range. This method uses the calculable radiation of a blackbody radiator and optical bandpass filters. The obtained results with relative standard measurement uncertainties between 5 % and 19 % are consistent with previous calibrations at PTB's national primary detector standard and extend the usable wavelength range.