Articles | Volume 8, issue 1
https://doi.org/10.5194/jsss-8-207-2019
https://doi.org/10.5194/jsss-8-207-2019
Regular research article
 | 
17 May 2019
Regular research article |  | 17 May 2019

Double entry method for the verification of data a chromatography data system receives

David Thomas Marehn, Detlef Wilhelm, Heike Pospisil, and Roberto Pizzoferrato

Viewed

Total article views: 3,971 (including HTML, PDF, and XML)
HTML PDF XML Total Supplement BibTeX EndNote
3,443 477 51 3,971 195 52 49
  • HTML: 3,443
  • PDF: 477
  • XML: 51
  • Total: 3,971
  • Supplement: 195
  • BibTeX: 52
  • EndNote: 49
Views and downloads (calculated since 17 May 2019)
Cumulative views and downloads (calculated since 17 May 2019)

Viewed (geographical distribution)

Total article views: 3,103 (including HTML, PDF, and XML) Thereof 2,935 with geography defined and 168 with unknown origin.
Country # Views %
  • 1
1
 
 
 
 
Latest update: 14 Dec 2024
Download
Short summary
In the field of pharmacy using diagnostic techniques like liquid chromatography a correct data acquisition process has to be guaranteed. So it is necessary to verify provided data of the device using two software instances collecting the generated raw data synchronously as we have done by a self-written instance. That way we found modifications on incoming raw data like time shifts and not expected interpolations during the storage procedure of one of two tested chromatography software packages.