Articles | Volume 11, issue 1
https://doi.org/10.5194/jsss-11-61-2022
https://doi.org/10.5194/jsss-11-61-2022
Regular research article
 | 
23 Feb 2022
Regular research article |  | 23 Feb 2022

Particularities of pyroelectric detectors in absolute measurements of chopped radiation shown for the example of a spectral responsivity calibration in the near- and mid-infrared spectral range at two primary radiometric standards

Tobias Pohl, Peter Meindl, Jörg Hollandt, Uwe Johannsen, and Lutz Werner

Viewed

Total article views: 1,035 (including HTML, PDF, and XML)
HTML PDF XML Total BibTeX EndNote
551 449 35 1,035 27 25
  • HTML: 551
  • PDF: 449
  • XML: 35
  • Total: 1,035
  • BibTeX: 27
  • EndNote: 25
Views and downloads (calculated since 23 Feb 2022)
Cumulative views and downloads (calculated since 23 Feb 2022)

Viewed (geographical distribution)

Total article views: 998 (including HTML, PDF, and XML) Thereof 998 with geography defined and 0 with unknown origin.
Country # Views %
  • 1
1
 
 
 
 

Cited

Latest update: 07 Nov 2024
Download
Short summary
The Physikalisch-Technische Bundesanstalt calibrated pyroelectric detectors at two independent primary radiometric standards regarding their spectral responsivity. The SI traceable measurements in the spectral range between 1.5 and 14 µm are consistent within the measurement uncertainty in the range of 1 % and 14 %. The importance of the correct read-out of lock-in amplifiers and the accurate consideration of the temporal shape of the chopped radiant flux behind the chopper wheel are shown.