AC characteristics of low-ohmic foil shunts influenced by eddy currents in the mounting body

Mario Schönecker-Baußmann

Model code and software

Supplementary dataset for Journal Article "AC Characteristics of Low Ohmic Foil Shunts Influenced by Eddy Currents in the Mounting Body" M. Schönecker-Baußmann https://doi.org/10.5281/zenodo.3369544

Fast Henry User's Guide M. Kamon, L. Silveira, C. Smithhisler, and J. White https://github.com/ediloren/FastHenry2/tree/WRCad

GNU Octave J. W. Eaton, D. Bateman, S. Hauberg, and R. Wehbring http://www.gnu.org/software/octave/octave.pdf

Download
Short summary
We came across some problems with the current measurement shunts while building a transfer normal power analyzer for 150 kHz within a cooperation project of the manufacturer ZES ZIMMER along with PTB, the National Metrology Institute of Germany and Bundesnetzagentur Berlin. We decided to utilize simulations with the numerical field simulation program Fast Henry to determine the cause of this frequency behavior. We found adequate justification and give recommendations for the shunt manufacturing.