Bilateral comparison for determining the dynamic characteristic values of contact thermometers in fluids
Silke Augustin
CORRESPONDING AUTHOR
Institute of Process Measurement and Sensor Technology, TU Ilmenau,
98693 Ilmenau, Germany
Thomas Fröhlich
Institute of Process Measurement and Sensor Technology, TU Ilmenau,
98693 Ilmenau, Germany
Marc Schalles
Institute of Process Measurement and Sensor Technology, TU Ilmenau,
98693 Ilmenau, Germany
Stefan Krummeck
JUMO GmbH & Co. KG, 36039 Fulda, Germany
Related authors
Silke Augustin and Marc Schalles
J. Sens. Sens. Syst., 12, 197–204, https://doi.org/10.5194/jsss-12-197-2023, https://doi.org/10.5194/jsss-12-197-2023, 2023
Short summary
Short summary
Temperature measurements are of great importance for the control of processes, for quality assurance, and for plant safety. It is not always possible to measure temperatures directly in the medium but on the surface of vessels and pipelines, etc. However, when determining the surface temperature with contact thermometers, a number of influencing variables, some of which are difficult to determine, affect accurate and reproducible temperature measurement. These are presented in the article.
Silke Augustin and Marc Schalles
J. Sens. Sens. Syst., 12, 197–204, https://doi.org/10.5194/jsss-12-197-2023, https://doi.org/10.5194/jsss-12-197-2023, 2023
Short summary
Short summary
Temperature measurements are of great importance for the control of processes, for quality assurance, and for plant safety. It is not always possible to measure temperatures directly in the medium but on the surface of vessels and pipelines, etc. However, when determining the surface temperature with contact thermometers, a number of influencing variables, some of which are difficult to determine, affect accurate and reproducible temperature measurement. These are presented in the article.
Markus Pabst, Maximilian Darnieder, René Theska, and Thomas Fröhlich
J. Sens. Sens. Syst., 11, 109–116, https://doi.org/10.5194/jsss-11-109-2022, https://doi.org/10.5194/jsss-11-109-2022, 2022
Short summary
Short summary
An adjustment concept for high-precision weighing cells has been developed and experimentally investigated. It allows for minimization of uncertainty contributions due to tilt while maximizing sensitivity to mass. By applying the adjustment capabilities to new weighing cells, the devices can simultaneously become more accurate, more adaptable and more robust.
Maximilian Darnieder, Markus Pabst, Ronny Wenig, Lena Zentner, René Theska, and Thomas Fröhlich
J. Sens. Sens. Syst., 7, 587–600, https://doi.org/10.5194/jsss-7-587-2018, https://doi.org/10.5194/jsss-7-587-2018, 2018
Related subject area
Measurement theory, uncertainty and modeling of measurements: Measurement uncertainty
Metrological assessment of a robotic total station for use in post-earthquake emergency conditions
Approximate sequential Bayesian filtering to estimate 222Rn emanation from 226Ra sources using spectral time series
Evaluation of precision, accuracy and threshold for the design of vibrotactile feedback in eye tracking applications
Influence of measurement uncertainty on machine learning results demonstrated for a smart gas sensor
Towards efficient application-dependent dimensional measurements with computed tomography: optimized reduction of measurement duration using continuous scan mode: experimental investigations
Structure of digital metrological twins as software for uncertainty estimation
Assessment of uncertainties for measurements of total near-normal emissivity of low-emissivity foils with an industrial emissometer
Measurement uncertainty assessment for virtual assembly
Measurements at laser materials processing machines: spectrum deconvolution including uncertainties and model selection
Determination of the single point precision associated with tactile gear measurements in scanning mode
Determination of the single point uncertainty of customized polymer gear wheels using structured-light scanning with various polygonization settings
Uncertainty evaluation in industrial pressure measurement
Amplitude–phase calibration of tri-axial accelerometers in the low-frequency range by a LDV
Calibration of torque measurement under constant rotation in a wind turbine test bench
Modeling the unified measurement uncertainty of deflectometric and plenoptic 3-D sensors
Sensors 4.0 – smart sensors and measurement technology enable Industry 4.0
Test equipment and its effect on the calibration of instrument transformers
A pathway to eliminate the gas flow dependency of a hydrocarbon sensor for automotive exhaust applications
Statistical characterization of evaluation strategies for fringe projection systems by means of a model-based noise prediction
Evaluation of dynamic measurement uncertainty – an open-source software package to bridge theory and practice
Towards assessing online uncertainty for three-phase flow metering in the oil and gas industry
Giulio D'Emilia and Emanuela Natale
J. Sens. Sens. Syst., 12, 187–195, https://doi.org/10.5194/jsss-12-187-2023, https://doi.org/10.5194/jsss-12-187-2023, 2023
Short summary
Short summary
After an earthquake, many resources are spent to evaluate the level of damage suffered by buildings and infrastructure. To improve the efficiency and reduce the subjectivity of such assessments, a robotic total station (RTS) can be a useful support. This work aims evaluates the performance of a RTS for use in emergency conditions. Furthermore, the analysis of reduced configurations of the monitoring points highlights interesting aspects for defining a standardized, simplified procedure.
Florian Mertes, Stefan Röttger, and Annette Röttger
J. Sens. Sens. Syst., 12, 147–161, https://doi.org/10.5194/jsss-12-147-2023, https://doi.org/10.5194/jsss-12-147-2023, 2023
Short summary
Short summary
In this work, a novel approach to deduce the release of the natural radioactive noble gas 222Rn from solid sources containing the isotope 226Ra is presented. Therein, supporting radioactivity measurements of the source are used in conjunction with a theoretical description of the dynamics. For radiation protection and environmental research, reliable and comparable 222Rn measurements, and therefore reference atmospheres of 222Rn, are needed. This work improves their realization.
Anke Fischer, Thomas M. Wendt, Lukas Stiglmeier, Philipp Gawron, and Kristof Van Laerhoven
J. Sens. Sens. Syst., 12, 103–109, https://doi.org/10.5194/jsss-12-103-2023, https://doi.org/10.5194/jsss-12-103-2023, 2023
Short summary
Short summary
In this work, we investigated whether artificial feedback in the form of vibration has an effect on the accuracy and precision of eye tracking glasses. The results enabled the development of design proposals for vibrotactile feedback in gaze-based control systems for robotic arms.
Tanja Dorst, Tizian Schneider, Sascha Eichstädt, and Andreas Schütze
J. Sens. Sens. Syst., 12, 45–60, https://doi.org/10.5194/jsss-12-45-2023, https://doi.org/10.5194/jsss-12-45-2023, 2023
Short summary
Short summary
A fundamental problem of machine learning (ML) is measurement uncertainty and the influence on ML results. Measurement uncertainty, which is critical in hazardous gas detection, is directly addressed in this paper. A previously published toolbox is extended for regression. One of the benefits of this approach is obtaining a better understanding of where the overall system should be improved. This can be achieved by either improving the trained ML model or using a sensor with higher precision.
Christian Orgeldinger, Florian Wohlgemuth, Andreas Michael Müller, and Tino Hausotte
J. Sens. Sens. Syst., 11, 219–223, https://doi.org/10.5194/jsss-11-219-2022, https://doi.org/10.5194/jsss-11-219-2022, 2022
Short summary
Short summary
Industrial X-ray computed tomography is a holistic measurement technique for dimensional metrology. However, the relatively long measurement time is a hindrance to its application. A possibility to reduce measurement times, the continuous scan mode, is characterized in this work. The question is how much the time reduction impacts the accuracy of the dimensional measurements. The core result of the paper is an estimate of the effect along with experimental proof that this estimate is reasonable.
Ivan Poroskun, Christian Rothleitner, and Daniel Heißelmann
J. Sens. Sens. Syst., 11, 75–82, https://doi.org/10.5194/jsss-11-75-2022, https://doi.org/10.5194/jsss-11-75-2022, 2022
Short summary
Short summary
The paper proposes a structure for the creation of new simulation software for uncertainty estimation. The structure was derived from the well-established VCMM. To make it easy to apply the software structure to specific projects, a supporting software library (written in C++) was created. The library provides all the components necessary to create software for uncertainty estimation. The software structure and library proposed can be used in different domains of metrology.
Jacques Hameury, Guillaume Failleau, Mariacarla Arduini, Jochen Manara, Elena Kononogova, Albert Adibekyan, Christian Monte, Alexander Kirmes, Eric Palacio, and Holger Simon
J. Sens. Sens. Syst., 10, 135–152, https://doi.org/10.5194/jsss-10-135-2021, https://doi.org/10.5194/jsss-10-135-2021, 2021
Short summary
Short summary
A TIR100-2 emissometer is used to measure the emissivity of thermal insulation products using the
low-emissivity effectto increase the thermal resistance, but the uncertainties of measurements on low-emissivity foils were not known. The study allowed the determination of the uncertainty through the detailed analysis of the measurement principle and the determination of the influences of the parameters. The measurement results were validated by comparison to a reference measurement setup.
Manuel Kaufmann, Ira Effenberger, and Marco F. Huber
J. Sens. Sens. Syst., 10, 101–108, https://doi.org/10.5194/jsss-10-101-2021, https://doi.org/10.5194/jsss-10-101-2021, 2021
Short summary
Short summary
Virtual assembly (VA) is a method for the quality prediction of assemblies considering local form deviations of relevant geometries. Point clouds of measured objects are registered in order to recreate the objects’ hypothetical physical assembly state, which is strongly influenced by the measurement uncertainty of individual points. Thus, we studied the propagation of uncertainties by VA. The results reveal larger propagated uncertainties by VA compared to the unconstrained Gaussian best fit.
Rolf Behrens, Björn Pullner, and Marcel Reginatto
J. Sens. Sens. Syst., 10, 13–18, https://doi.org/10.5194/jsss-10-13-2021, https://doi.org/10.5194/jsss-10-13-2021, 2021
Short summary
Short summary
The dose rate and spectral distribution of X-ray emissions from laser materials processing have been determined using a thermoluminescence detector (TLD)-based spectrometer. The penetration depth of the radiation into the spectrometer depends on its energy, so that the energy-resolved spectrum of the radiation can be calculated from the TLD dose values by means of mathematical methods (Bayesian deconvolution). The measurements are traceable to the SI.
Andreas Michael Müller and Tino Hausotte
J. Sens. Sens. Syst., 9, 61–70, https://doi.org/10.5194/jsss-9-61-2020, https://doi.org/10.5194/jsss-9-61-2020, 2020
Short summary
Short summary
The framework of the single point uncertainty presents a methodology to determine the local measurement uncertainty for a measurement setup. The targets of the investigation were spur (involute) steel gear wheels using a CMM in scanning mode in combination with a rotatory table, as well as a single scan of the complete gear profile without the use of a rotatory table, using the
free-form scanCMM functionality. Both methods were examined with respect to their obtained single point uncertainty.
Andreas Michael Müller, Dominik Schubert, Dietmar Drummer, and Tino Hausotte
J. Sens. Sens. Syst., 9, 51–60, https://doi.org/10.5194/jsss-9-51-2020, https://doi.org/10.5194/jsss-9-51-2020, 2020
Short summary
Short summary
This paper aims to demonstrate the complete workflow of the determination of the local measurement uncertainty and its components (systematic and random measurement error) for a given measurement task. It was shown for an optical measurement setup in combination with an industrial X-ray computed tomography reference measurement system that different necessary colouring methods of polymer gear wheels have a measurable influence on the local distribution of the measurement uncertainty.
Nadine Schiering and Olaf Schnelle-Werner
J. Sens. Sens. Syst., 8, 251–259, https://doi.org/10.5194/jsss-8-251-2019, https://doi.org/10.5194/jsss-8-251-2019, 2019
Short summary
Short summary
When using pressure gauges in industry the uncertainty contribution due to the calibration should be expanded by the uncertainty contributions due to specific application. There the end user must investigate these sources or other additional influences. In addition, the new challenges in terms of calibration and traceability when applying Smart Sensor Technologies and Digitally Networked Measurement Systems were addressed.
Giulio D'Emilia, Antonella Gaspari, and Emanuela Natale
J. Sens. Sens. Syst., 8, 223–231, https://doi.org/10.5194/jsss-8-223-2019, https://doi.org/10.5194/jsss-8-223-2019, 2019
Short summary
Short summary
A method for amplitude–phase calibration of tri-axial accelerometers in the low-frequency range is proposed, based on a linear slide, used to excite all the axes of the accelerometer at the same time, and a laser Doppler vibrometer (LDV) as a reference. Results show that the phase is a critical aspect to consider in calibration, more than the amplitude, and the comparison with the theoretical model is useful to verify the hypotheses. Different behaviours result, depending on the measuring chain.
Paula Weidinger, Gisa Foyer, Stefan Kock, Jonas Gnauert, and Rolf Kumme
J. Sens. Sens. Syst., 8, 149–159, https://doi.org/10.5194/jsss-8-149-2019, https://doi.org/10.5194/jsss-8-149-2019, 2019
Short summary
Short summary
To determine the efficiency of multi-MW wind turbines, the torque measurement in wind turbine test benches has to be performed with a high accuracy. To this end, the torque measurement has to be traced to national standards. This can be done by using a novel torque transfer standard in combination with a newly developed torque calibration procedure under constant rotation. The calibration procedure was performed on a wind turbine test bench.
Mathias Ziebarth, Niclas Zeller, Michael Heizmann, and Franz Quint
J. Sens. Sens. Syst., 7, 517–533, https://doi.org/10.5194/jsss-7-517-2018, https://doi.org/10.5194/jsss-7-517-2018, 2018
Short summary
Short summary
We propose a measurement uncertainty model of two complementary optical sensors: a deflectometric and a plenoptic sensor. The deflectometric sensor uses active triangulation and works best on specular surfaces, while the plenoptic sensor uses passive triangulation and works best on textured, diffusely reflecting surfaces. The predicted uncertainties can be used to obtain optimized measurements for varying surface properties. The models are validated exemplarily based on real measurements.
Andreas Schütze, Nikolai Helwig, and Tizian Schneider
J. Sens. Sens. Syst., 7, 359–371, https://doi.org/10.5194/jsss-7-359-2018, https://doi.org/10.5194/jsss-7-359-2018, 2018
Short summary
Short summary
“Industrie 4.0” or the Industrial Internet of Things (IIoT) describe the current (r)evolution in industrial automation and control. This is fundamentally based on smart sensors, which generate data and allow further functionality from self-monitoring and self-configuration to condition monitoring of complex processes. The paper reviews the development of sensor technology over the last 2 centuries and highlights some of the potential that can be achieved with smart sensors and data analysis.
Enrico Mohns and Peter Räther
J. Sens. Sens. Syst., 7, 339–347, https://doi.org/10.5194/jsss-7-339-2018, https://doi.org/10.5194/jsss-7-339-2018, 2018
Short summary
Short summary
Test centres supplying accuracy tests for instrument transformers must provide measurement uncertainties for their quality management. In this work, ratio error and phase displacement of instrument transformers are discussed. The traceability to the national standards of PTB, the attainable uncertainty and the permitted error limits of test equipment for testing instrument transformers are presented.
An example of an uncertainty budget for a current transformer of the class 0,2 S is given.
Gunter Hagen, Antonia Harsch, and Ralf Moos
J. Sens. Sens. Syst., 7, 79–84, https://doi.org/10.5194/jsss-7-79-2018, https://doi.org/10.5194/jsss-7-79-2018, 2018
Short summary
Short summary
Monitoring hydrocarbon concentrations in automotive exhausts is affected by flow rate changes. The signal of thermoelectric gas sensors is a thermovoltage. Its origin is a temperature difference that depends on the flow rate. To avoid this noise effect, the sensor can be installed in a defined bypass position. As shown by simulation and experiments, the gas flow around the sensor is almost turbulence-free and the signal only depends on the hydrocarbon concentration and not on the flow rate.
Marc Fischer, Marcus Petz, and Rainer Tutsch
J. Sens. Sens. Syst., 6, 145–153, https://doi.org/10.5194/jsss-6-145-2017, https://doi.org/10.5194/jsss-6-145-2017, 2017
Short summary
Short summary
For optical measurement systems, camera noise is the dominant uncertainty factor when optically cooperative surfaces are measured in a stable environment. In this work it will be shown that the resulting spatial noise can be estimated by means of a noise model. The capability of this approach will be demonstrated for a fringe projection system. This provides a valuable tool for a statistical comparison of different evaluation strategies, which is shown for two different triangulation procedures.
Sascha Eichstädt, Clemens Elster, Ian M. Smith, and Trevor J. Esward
J. Sens. Sens. Syst., 6, 97–105, https://doi.org/10.5194/jsss-6-97-2017, https://doi.org/10.5194/jsss-6-97-2017, 2017
Short summary
Short summary
The challenges in the analysis of dynamic measurements significantly limit the use of existing calibration facilities and mathematical methodologies. Several international research projects have developed methodologies for the treatment of dynamic measurements at the NMI level. The transfer to industry, though, remains challenging. Therefore, this work provides a link from these activities to industrial end-users by means of the new versatile and comprehensive software package
PyDynamic.
M. P. Henry, M. S. Tombs, and F. B. Zhou
J. Sens. Sens. Syst., 3, 97–103, https://doi.org/10.5194/jsss-3-97-2014, https://doi.org/10.5194/jsss-3-97-2014, 2014
Cited articles
Augustin, S. and Fröhlich, T.:
Dynamisches Verhalten von Berührungsthermometern, tm-Tech. Mess.
83, 402–409, https://doi.org/10.1515/teme-2015-0068,
2017.
Augustin, S., Fröhlich, T., and Ament, C.:
Temperature-Dependent Dynamic Behaviour of Process Temperature Sensors,
International Journal of Thermophysics, Springer Verlag, https://doi.org/10.1007/s10765-015-1869-4, 2015.
Augustin, S., Fröhlich, T., and Heydrich, M.:
Bestimmung der Messunsicherheit dynamischer Kennwerte von
Berührungsthermometern in strömender Luft, tm-Tech. Mess., 84, 73–82, https://doi.org/10.1515/teme-2016-0055,
2017.
Bernhard, F.: Handbuch der Technischen Temperaturmessung, 2.
Auflage, Springer-Verlag, 2014.
DAkkS: DAkkS-DKD-R 5-3: Kalibrierung von Thermoelementen,
Beuth-Verlag, 1. Neuauflage, 2010a.
DAkkS: DAkkS-DKD-R 5-1: Kalibrierung von Widerstandsthermometern,
Beuth-Verlag, 1. Neuauflage, 2010b.
DIN: DIN EN 60751: Industrielle Platin-Widerstandsthermometer und
Platin-Temperatursensoren:2008, Beuth Verlag GmbH, 2010.
DIN: DIN EN 61515:2016: Mineralisolierte metallgeschirmte
Mantelthermoelementleitungen und Mantelthermoelemente, Beuth Verlag GmbH,
2017.
Fröhlich, T. and Blumröder,
G.: Bestimmung des Verhältnisses der Zeitkonstanten zweier
Trägheitsglieder erster Ordnung aus der Differenz der Sprungantworten,
Technisches Messen, Band 65, 1998.
JUMO: Prüfverfahren und Messablauf zur Ermittlung der
Ansprechzeiten, Internes Papier, JUMO, Fulda, 1994.
Lieneweg, F.: Übergangsfunktion (Anzeigeverzögerung)
von Thermometern – Aufnahmetechnik, Meßergebnisse, Auswertungen, Archiv
für Technisches Messen, 1964, R46–R53, 1964.
Mammen, H., Krapf, G., Hoffmann, C., Sasiuk, T., Pufke, M., Marin, S., Fröhlich, T.: Prüfeinrichtung zur Untersuchung
des dynamischen Verhaltens von Berührungsthermometern in Wasser,
TEMPERATUR 2017, PTB Berlin, Tagungsband, 163–168, 17 and 18 Mai 2017,
Berlin, 2017.
VDI: VDI-Wärmeatlas, 11. bearbeitete und erweiterte Auflage,
Springer-Verlag Berlin Heidelberg, 2013.
VDI/VDE: VDI/VDE-Richtlinie 3522: Dynamisches Verhalten von
Berührungsthermometern, VDI-Verlag GmbH, 2014.
Wienecke, J.: Dynamik-Prüfeinrichtung für
industrielle Temperaturfühler, Diplomarbeit, TU Ilmenau, 1993.
Short summary
In the data sheets of the thermometer manufacturers, different specifications can be found to describe the dynamic properties of contact thermometers. The present paper describes the results of a bilateral comparison made for the first time for determining dynamic parameters for two thermocouples in the laboratories of the JUMO GmbH & Co. KG Fulda company and at the Institute of Process Measurement and Sensor Technology of the TU Ilmenau with similar facilities, but different results.
In the data sheets of the thermometer manufacturers, different specifications can be found to...
Special issue